Performance Considerations
70
Anemia
Anemia causes decreased arterial oxygen content. Although SpO
2
readings may appear normal, an anemic patient may be hypoxic.
Correcting anemia can improve arterial oxygen content. The pulse
oximeter may fail to provide an SpO
2
if hemoglobin levels fall below
5 gm/dl.
Saturation
The N
-
560 will only measure saturation levels between 1 and 100%.
Pulse rates
The N
-
560 will only measure pulse rates between 20 and 250 beats per
minute. Detected pulse rates outside the range of 20 to 250 beats per
minute are displayed as the closest value within the range.
Sensor Performance Considerations
WARNING: Pulse oximetry readings and pulse signal can be affected
by certain ambient conditions, sensor application errors, and certain
patient conditions.
Inaccurate measurements can be caused by:
•
incorrect application of the sensor
•
placement of the sensor on an extremity with a blood pressure
cuff, arterial catheter, or intravascular line
•
ambient light
•
excessive patient movement
Loss
-
of
-
pulse signal can occur for the following reasons:
•
the sensor is applied too tightly
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Страница 12: ...Introduction 4...
Страница 22: ...Description of Controls Indicators and Symbols 14...
Страница 50: ...Using the N 560 42...
Страница 54: ...N 560 Trend 46...
Страница 70: ...Printing 62...
Страница 76: ...Sensors and Accessories 68...
Страница 82: ...Operator s Menu 74...
Страница 90: ...Troubleshooting 82...
Страница 98: ...Technical Information 90...
Страница 118: ...Specifications 110...
Страница 122: ...Index 114...
Страница 123: ......