Using the N-560
26
No Sensor Attached
When the sensor is not attached the N-560 displays dashes (---) and the
Pulse Search
indicator is not lit, indicating that the N-560 failed to detect
a sensor.
Sensor Message
The Sensor Message feature is an indication that the sensor position or
site needs to be considered. The
Sensor Message
indicator lights when
the N-560 cannot determine an SpO
2
level or a pulse rate. The Sensor
Message recommendations for improving the signal are:
•
Reposition sensor
•
Check or change adhesive wrap
•
Choose alternate site
•
Warm site
•
Cover sensor
•
Use forehead, nasal, or ear sensor (adult patients only)
•
Use
O
XI
M
AX
adhesive sensor
•
Secure cable
•
Secure with headband (
M
AX
-
F
AST
)
•
Remove nail polish
•
Loosen sensor (too tight)
•
Isolate external interference (electrosurgical device, cell phone)
•
Clean site (MAX
-
R)