Printing
N-560 Operator’s Manual
55
P r i n t i n g
Printing N-560 Real-Time Data
Real-time data is continuously sent to the data port on the back of the
N
-
560. Patient data can be obtained through the data port by connecting
to a computer or serial printer. When a real-time printout or display is
being transmitted to a printer or computer, a new line is printed/displayed
every 2 seconds. Column headings are printed/displayed every 25 lines,
or if one of the values in the column heading is changed.
Note: If the data output stops transmitting, the N
-
560 must be turned off
and then turned back on.
When connecting the N
-
560 to a printer or computer, verify proper
operation before clinical use. Both the N
-
560 and the printer or PC must
be connected to a grounded AC outlet.
Any printer or computer connected to the N-560's data port must be
certified according to IEC Standard 60950. All combinations of
equipment must be in compliance with IEC Standard 60601
-
1
-
1 systems
requirements. Anyone who connects a printer or PC to the data output
port configures a medical system and is therefore responsible that the
system complies with the requirements of system standard IEC Standard
60601
-
1
-
1 and the electromagnetic compatibility system standard IEC
Standard 60601
-
1
-
2.
1. Data Port Connector
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Страница 12: ...Introduction 4...
Страница 22: ...Description of Controls Indicators and Symbols 14...
Страница 50: ...Using the N 560 42...
Страница 54: ...N 560 Trend 46...
Страница 70: ...Printing 62...
Страница 76: ...Sensors and Accessories 68...
Страница 82: ...Operator s Menu 74...
Страница 90: ...Troubleshooting 82...
Страница 98: ...Technical Information 90...
Страница 118: ...Specifications 110...
Страница 122: ...Index 114...
Страница 123: ......