Specifications
N-560 Operator’s Manual
107
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Complies
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the AC mains voltage prior to application of the test
level.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHZ to
80 MHz
3 Vrms
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
N-560, including cables,
than the recommended
separation distance
calculated from the
equation applicable to the
frequency of the
transmitter.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Table 9: Electromagnetic Immunity (Cont.)
The N-560 is intended for use in the electromagnetic environment specified
below. The customer or user of the N-560 should assure that it is used in such
an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment — Guidance
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Страница 12: ...Introduction 4...
Страница 22: ...Description of Controls Indicators and Symbols 14...
Страница 50: ...Using the N 560 42...
Страница 54: ...N 560 Trend 46...
Страница 70: ...Printing 62...
Страница 76: ...Sensors and Accessories 68...
Страница 82: ...Operator s Menu 74...
Страница 90: ...Troubleshooting 82...
Страница 98: ...Technical Information 90...
Страница 118: ...Specifications 110...
Страница 122: ...Index 114...
Страница 123: ......