![Nellcor OXIMAX N-560 Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_operators-manual_3908427056.webp)
Using the Data Port
48
No hardware flow control is used. However, in the ASCII mode, XON/
XOFF flow control is supported.
Data Port Pinouts
TXD represents the Transmit Data line, and RXD is the Receive Data
line.
The pinouts for the data port are listed in Table 1.
The pin layouts (as viewed from the rear panel of the N
-
560) are
illustrated in Figure 3. The conductive shell is connected to earth ground
when connected to a PC or printer.
Table 1: Data Port Pinouts
Pin
Signal Name
1 RXD+ (RS-422 positive input)
2
RXD (RS
-
232 input)
3
TXD (RS
-
232 output)
4
TXD+ (RS-422 positive output))
5
Signal Ground (isolated from Earth Ground)
6
NC (No connection)
7
Normally Open (relay closure nurse call, normally open)
8
Normally Closed (relay closure nurse call, normally closed)
9
RXD- (RS-422 negative output)
10
Signal Ground (isolated from earth ground)
11
Nurse Call (RS-232 level output)
12
TXD- (RS-422 negative output)
13
NC
14
NC
15
Nurse Call Common
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Страница 12: ...Introduction 4...
Страница 22: ...Description of Controls Indicators and Symbols 14...
Страница 50: ...Using the N 560 42...
Страница 54: ...N 560 Trend 46...
Страница 70: ...Printing 62...
Страница 76: ...Sensors and Accessories 68...
Страница 82: ...Operator s Menu 74...
Страница 90: ...Troubleshooting 82...
Страница 98: ...Technical Information 90...
Страница 118: ...Specifications 110...
Страница 122: ...Index 114...
Страница 123: ......