13
Color Menu
Gamma:
Press
◄
or
►
to select a preset
Gamma
(
1.8
,
2.0
,
2.2
,
2.4
or
2.6
),
Video
or
PACS
Notes:
1.
Video
is a color corrected Look Up Table (LUT).
2.
PACS
is a DICOM-like LUT.
Note:
This product is capable of displaying Radiology (PACS) images for
reference, not diagnostic, purposes only.
Color Temperature:
Press the
◄
or
►
button to select a preset color temperature.
Note:
When a preset is selected and any
parameter is changed, the current values are copied to the User presets and
User
is selected.
Red, Green, Blue:
Press the
◄
or
►
button to increase or decrease the intensity of the selected color.
Saturation:
Press
◄
or
►
to set the saturation (color intensity) of the image.
Hue:
Press
◄
or
►
to set the hue (color tint) of the image.
Brightness:
Press the
◄
or
►
button to adjust brightness. This is a duplicate of the Brightness control described on
page 5.
Contrast:
Press the
◄
or
►
button to adjust contrast. This is a duplicate of the Contrast control described on page 5.
Video Level:
Normal:
Provides a dynamic range equivalent to the incoming signal with no change.
Expanded:
Expands the signal level such that an input signal with range 16 (black) to 235 (white) will be
expanded to a range of 0 (black) to 255 (white).
Note:
If a signal is already full range (0 to 255), changing
the setting to
Expanded
will cause the signal to “clip” or “saturate”.
Press the
◄
or
►
button to select.
Содержание Radiance G3
Страница 1: ...Radiance G3 26 USER MANUAL ENGLISH ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 25: ...17 ...
Страница 29: ...21 ...
Страница 32: ...24 ...
Страница 33: ...25 ...
Страница 34: ...26 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 ...
Страница 35: ......