OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
8-1
S e c t i o n
8
Specifications
Intended Use
Olympic Driers are intended to rapidly dry
pasteurized medical equipment, primarily
respiratory therapy and anesthesia parts. The Drier
should be used after disinfection by pasteurization
or other liquid chemical means.
Dimensions
Size
Model 43:
26 in. wide x 20.5 in. deep x 86 in. high
66 cm wide x 52 cm deep x 218 cm high
Model 44:
46.5 in. wide x 20.5 in. deep x 86 in. high
118 cm wide x 52 cm deep x 218 cm high
Weight
Model 43:
125 lb
57 kg
Model 44:
247 lb
112 kg
Electrical
Power Requirements
Model 43 and Model 44 (100V)
Voltage:
100V~, single phase
Consumption:
9.5A (Model 43)
18.0A (Model 44)
Frequency:50/60 Hz
Fuse:
15A (Model 43)
25A (Model 44)
Model 43 and Model 44 (120V)
Voltage:
120V~, single phase
Consumption:
8.0A (Model 43)
16.0A (Model 44)
Frequency:60 Hz
Fuse:
12A (Model 43)
20A (Model 44)
Model 43 and Model 44 (220-230V)*
Voltage:
220-230V~, single phase
Consumption:
4.5A (Model 43)
9.0A (Model 44)
Frequency:50/60 Hz
Fuse:
6A (Model 43)
12A (Model 44)
*CE mark applies to 220-230V devices.
Ground
Does not exceed 0.13
Ω
Fuses
Two internal 250V FLM time-delay fuses
Environmental
Temperature
Operating: 65–85°F
18–30°C
Shipping: -40–158°F
-40–70°C
Storage:
-40–158°F
-40–70°C
Relative Humidity (RH)
Operating: 0–95%, non-condensing
Non-operating: 0–100%, condensing
Altitude (Maximum)
Up to 6560 ft
Up to 2000 m
Operating and Non-Operating Pressure
0.015–0.0295 hPa
0.5–1.0 inHg
Noise
Maximum noise level: less than 65dB(A) from six
feet away
Содержание OLYMPIC STERILE-DRIER 43
Страница 2: ......
Страница 4: ......
Страница 10: ...1 4 OLYMPIC STERILE DRIER Model 43 44 Instruction Manual Accessories ...
Страница 18: ...2 8 OLYMPIC STERILE DRIER Model 43 44 Instruction Manual Testing the Drier ...
Страница 22: ...3 4 OLYMPIC STERILE DRIER Model 43 44 Instruction Manual Operating the Drier ...