OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
4-1
S e c t i o n
4
Maintenance
Weekly Maintenance
Cleaning the Drier
Required Items:
Soft cleaning cloths
Mild cleaning detergent (e.g., Virex
TM
Tb, Virustat
®
, Coverage
®
) or
mild soap-and-water solution
Disinfectant (e.g., Sporicidin
®
)
To clean the Drier:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Using a soft cloth dampened with mild detergent, wipe the exterior of the Drier,
including its power cord. Use a second cloth to dry all exterior surfaces.
WARNING
Electrical shock hazard when the enclosure is open. Unplug the power cord from the
electrical outlet before cleaning this device.
CAUTION
Only qualified technicians should perform maintenance procedures.
Never use alcohol to clean the Plexiglas door as alcohol will damage the plastic.
Do not allow liquid to enter the louver prefilter cover to avoid damaging the internal
filter system.
NOTE
#
To prevent contamination, perform cleanup procedures after each use.
Содержание OLYMPIC STERILE-DRIER 43
Страница 2: ......
Страница 4: ......
Страница 10: ...1 4 OLYMPIC STERILE DRIER Model 43 44 Instruction Manual Accessories ...
Страница 18: ...2 8 OLYMPIC STERILE DRIER Model 43 44 Instruction Manual Testing the Drier ...
Страница 22: ...3 4 OLYMPIC STERILE DRIER Model 43 44 Instruction Manual Operating the Drier ...