OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
6-5
Repair Procedures
5
From the side access panel, locate and remove the kep nuts (x2) that secure the
front control access panel assembly to the front column (see Figure 6.1). From
inside the blower access upper cabinet, remove the kep nut attached to the top
of the control panel.
6
To prevent ESD, attach an electrical grounding strap to yourself and to a
non-painted metallic surface in the upper cabinet of the Drier, such as where the
blower assemblies are located.
7
Pull out the panel assembly and disconnect the wires.
8
Remove the six nuts and plastic spacers from the control panel PCB.
To install the new control panel PCB:
1
Insert the six nuts and plastic spacers in the new control panel PCB.
2
Connect the wires to the panel assembly (see the wiring diagram on page A-1).
3
Secure the panel assembly with the kip nuts to the top and front of the
access panel.
4
Insert the front cover.
5
Tighten the screws inside the Drier door.
6
Attach the side access panel, then secure all the tray brackets and trays in
the Drier.
7
Plug in the power cord, then press the power switch to the
ON
(|) position.
8
Test the device for proper operation (see page 2-4).