1 Product Description
The neoBLUE
®
Phototherapy System consists of two products – the neoBLUE LED
Phototherapy light source (light) and the neoBLUE LED Phototherapy roll stand.
Intended Use
The neoBLUE LED Phototherapy System is intended for the treatment of
neonatal hyperbilirubinemia. The light can be used for infants in a bassinet,
incubator, open bed, or radiant warmer.
Important! Before assembling the neoBLUE light and administering
phototherapy, read all sections of this manual carefully. There are safety
considerations that should be read and understood before use.
Physical Characteristics
The neoBLUE LED Phototherapy System is a floor-standing, mobile
phototherapy light that delivers a narrow band of high-intensity blue light via
blue light emitting diodes (LEDs) to provide treatment for neonatal
hyperbilirubinemia.
Light Source
The light consists of a lightweight plastic light enclosure. When used with the
neoBLUE roll stand, the light can be tilted and adjusted both horizontally and
vertically on the roll stand assembly. The light enclosure can be tilted to
approximately 40° up from horizontal (the resting position). The light
enclosure height can be adjusted vertically along the roll stand post, as well
as horizontally out from the roll stand post (proximity adjustment) to aid in
positioning the light. To help position the light over the infant, a red target light
can be briefly illuminated using the right rocker switch on the front panel. The
light can be used for infants in a bassinet, incubator, open bed, or radiant
warmer.
The light can be used independently of the roll stand. The light enclosure can
be placed directly on an incubator with a flat-topped surface.
Important! Medical electrical equipment needs special precautions
regarding EMC and needs to be installed and put into service according to
the EMC information provided in this document.
Warning! Portable and mobile RF communications equipment can
affect medical electrical equipment
Warning! Placement directly on incubator: The enclosure can be
placed on flat surfaces only. Confirm all rubber feet are fully seated on the top
of the enclosure to prevent slippage. When placing light enclosure directly on
incubator, care must be taken to ensure a safe operating environment.
Secure power cord to minimize risk of tripping.
Warning! Placement directly on incubator: The use of skin-
controlled mode (patient servo) of the incubator or radiant warmer is
neoBLUE
®
LED Phototherapy System User Manual
1
PN 001364 Rev. I
Содержание neoBLUE