6  Routine Cleaning and Maintenance 

6.1  Checking the Light Intensity  

It is recommended that the intensity of the light be checked per your hospital 
protocol, or at least every six months to ensure proper light intensity. Have a 
qualified technician test the intensity level and readjust the intensity 
potentiometers to achieve the desired output, if required.  Checking intensity 
before each use is recommended; however, as LEDs have very gradual 
degradation over time, it is not mandatory.  

 

 

Note:  The intensity of the light is not adversely affected if a few LEDs burn 
out. The LEDs are installed in groups of six, so failure of a single LED 
normally causes six to fail. 

 

 

Caution:  Only qualified personnel should perform service and repair. Use 
extreme care when working with exposed circuitry. 

6.2  Adjusting the Light Intensity 

There are two intensity settings, high and low. The desired setting is selected 
using the left rocker switch on the front panel of the light. The light output was 
factory calibrated with the neoBLUE

®

 Radiometer to provide an initial 

intensity of 35 µW/cm

2

/nm at the high setting and 15 µW/cm

2

/nm at the low 

setting

 

at a distance of 12 inches (30.5 cm) from the light enclosure to the 

baby. The light output can also be adjusted using the two potentiometers 
(located on the side of the light enclosure) to accommodate different 
distances. 

 

Note:  The light will continue to operate with minimal degradation over time.  
Timing of potentiometer adjustments is per hospital protocol.  

Important!  The light was factory calibrated with the neoBLUE

®

 Radiometer 

to deliver intensive phototherapy at a distance of 12 inches (30.5 cm) from 
the baby. Refer to the Service Manual for information on adjusting the 
intensity if using the light at other distances, and for information on using 
other meters. 

6.3  Cleaning 

 

 

 

Warning!  Disconnect the light from AC power before cleaning. 

Remove dust from the exterior of the light with a soft brush or soft cloth 
dampened with water. Sponge away remaining debris with a mild solution of 
detergent and water, a non-caustic commercial cleaner, or hospital 
disinfectant. 

Clean the lens with a soft cloth dampened with water. If water alone is 
ineffective in removing fingerprints or other markings, use a mild solution of 
detergent and water, a non-caustic commercial cleaner, or hospital 
disinfectant. 

 

 

  Caution:  Observe the following precautions: 

neoBLUE

®

 LED Phototherapy System User Manual 

15 

PN 001364 Rev. I 

Содержание neoBLUE

Страница 1: ...User Manual...

Страница 2: ...nual may change without notice Natus Medical Incorporated 5900 First Avenue South Seattle WA 98108 USA Telephone 1 650 802 0400 Fax 1 650 802 0401 Technical Service 1 800 303 0306 Technical Service Fa...

Страница 3: ...ure and Roll Stand 9 3 2 Front Panel Controls 10 3 3 Rear Panel 10 4 Assembly and Operating Instructions 12 5 Troubleshooting Guide 14 6 Routine Cleaning and Maintenance 15 6 1 Checking the Light Inte...

Страница 4: ...ure height can be adjusted vertically along the roll stand post as well as horizontally out from the roll stand post proximity adjustment to aid in positioning the light To help position the light ove...

Страница 5: ...rption of light by bilirubin and is thus considered to be the most effective for the degradation of bilirubin Blue LEDs do not emit energy in the ultraviolet UV region of the spectrum so there is no U...

Страница 6: ...The light is mains power operated The power cord plugs into a receptacle at the power inlet at the back of the light enclosure There are no single use components for the light A power cord strain rel...

Страница 7: ...ions to ensure the safety of the patient and those near the instrument In addition please refer to your hospital policy and procedure for phototherapy administration Warning The neoBLUE device should...

Страница 8: ...ment of Light Enclosure When attaching the light to any floor stand other than the neoBLUE roll stand confirm weight capacity of stand Tippage test must be done with the light in the fully extended po...

Страница 9: ...y and the unit from incidental debris or fluids Warning Photosensitive drugs The light generated can degrade photosensitive medications Do not place or store any drugs near or in the illuminated area...

Страница 10: ...neoBLUE system or place anything on top of the light The roll stand and light are not designed to support additional equipment If other equipment must be used in conjunction with this product the equ...

Страница 11: ...nsult Instructions for Use Always protect the baby s eyes with eye patches or equivalent Authorized European Representative Manufacture Date Manufacturer WEEE Directive Waste from Electrical and Elect...

Страница 12: ...ts the baby and the unit from incidental debris or fluids Height Adjustment This knob allows you to adjust the height of the light enclosure First loosen the knob then adjust the height of the light e...

Страница 13: ...center the light over the infant press this switch to project a red light over the centrally illuminated area 3 3 Rear Panel Power cord attachment Timer Timer The neoBLUE light is equipped with a tim...

Страница 14: ...bottom front sides and back edges are passive air conduction vents If the fans cease to operate contact Natus Technical Service or your authorized service provider Caution To avoid overheating the li...

Страница 15: ...s for whom phototherapy has been prescribed 3 Shield infant s eyes with protective eye shields designed for use during phototherapy Biliband Eye Protectors Sizes Micro P N 900644 Premature P N 900643...

Страница 16: ...cm from the baby Refer to the service manual for information on adjusting the intensity if using the light at other distances 7 Select High or Low intensity setting as appropriate for the patient 8 Mo...

Страница 17: ...cian check the components and replace as necessary Some LEDs are not lit One LED may have burned out causing six LEDs to go off Have a qualified technician test the intensity level and readjust the in...

Страница 18: ...the low setting at a distance of 12 inches 30 5 cm from the light enclosure to the baby The light output can also be adjusted using the two potentiometers located on the side of the light enclosure to...

Страница 19: ...rect operating temperature the vent filter on the back of the light should be kept clear and free of dust As part of routine maintenance the vent filters should be cleaned every month Remove black fil...

Страница 20: ...e surface area for phototherapy The intensity of the light is inversely related to the distance from the light source to the baby The light output can be adjusted using the two potentiometers located...

Страница 21: ...lectrical Mains 85 264 V 47 to 63 Hz Maximum Rating 3A 100 240V 50 60 Hz Fuses 4A 100 120V 50 60 Hz 2A 200 240V 50 60 Hz Safety Leakage current 100 A Audible Noise 60 dB Dimensions Maximum Height 6 ft...

Страница 22: ...01 1 1988 A1 1991 A2 1995 IEC 60601 1 2005 UL 60601 1 2006 ES 60601 1 2012 CSA C22 2 601 1 M90 R2006 CSA C22 2 60601 1 08 R2013 IEC 60601 2 50 2000 IEC 60601 2 50 2009 IEC 60601 1 2 2007 LED PHOTOTHER...

Отзывы: