SCXI-1327 Installation Guide
Note
For correct operation with the SCXI-1121 module, you must bypass the attenuators
on the excitation channels.
Connecting Signals
Caution
Refer to the
Safety Information
section before removing equipment covers
or connecting or disconnecting any signal wires.
To connect the signal to the terminal block, perform the following steps,
referring to Figures 1 and 2 as necessary:
1.
Unscrew the top cover screws and remove the top cover.
2.
Loosen the strain-relief screws and remove the strain-relief bar.
3.
Prepare your signal wire by stripping the insulation no more than
7 mm.
Table 1.
SCXI-1327 High-Voltage Attenuator Configuration
Switch Position
Description
Enables the attenuator for a specific channel.
This is the default setting.
Bypasses the attenuator for a specific channel.
100:1
1:1
100:1
1:1