Specifications
Appendix A
SCXI-1120 User Manual
A-2
© National Instruments Corporation
Rise time
4 Hz
0.12 sec
10 kHz
70
µ
sec
Slew rate
0.15 V/
µ
sec
Noise (400 kHz bandwidth)
1
Input (gain = 1,000)
4 Hz filter
100 nVrms
10 kHz filter
4
µ
Vrms
Output (gain = 1)
4 Hz
150
µ
Vrms
10 kHz
1 mVrms
Operating common-mode voltage
50 or 60 Hz
Channel to channel or
Channel to earth
250 Vrms
5
Common-mode rejection ratio,
50 or 60 Hz
1 k
Ω
in input leads
160 dB minimum at 4 Hz bandwidth
NMR (50 or 60 Hz)
60 dB at 4 Hz bandwidth
Input protection (continuous)
250 Vrms maximum
Output protection
Continuous short-to-ground
Power dissipation
7.5 W maximum
Cold-Junction Sensor
3
SCXI-1320
Accuracy
1.0
°
from 0
°
to 55
°
C
Output
10 mV/
°
C
SCXI-1328
Accuracy
4
0.35
°
from 15
°
to 35
°
C
0.65
°
from 0
°
to 15
°
and 35
°
to 55
°
C
Output
1.91 to 0.58 V from 0
°
to 55
°
C
1
Includes the combined effects of the SCXI-1120 and the AT-MIO-16F-5
2
Includes the combined effects of the SCXI-1120 and the AT-MIO-16X
3
Located on the SCXI-1320 and SCXI-1328 terminal blocks
4
Includes the combined effects of the temperature sensor accuracy and the temperature difference between the
temperature sensor and any screw terminal
5
Module tested using the UL 1244 standard to twice the working v 1,000 Vrms