Physical Characteristics
Weight (probe only)
96 g
Cable length
1.3 m
Calibration
This section contains calibration information and verification procedures for the
SA1000X.
Verification confirms whether the probe is operating within published
specifications. If the probe performs outside of the limits defined in the verification
procedures, return the probe to NI for repair.
Refer to
for more information on contacting NI.
Related tasks
■
Related reference
■
Test Equipment
Refer to
and
for a list of necessary equipment and, where applicable,
model recommendations for calibration of the SA1000X.
If you do not have the recommended equipment, select a substitute calibration
standard using the specifications listed in the Requirements column.
Equipment
Recommended
Model
Where Used
Requirements
2 × DMM
NI PXIe-4082
(× 2)
■
Measurement error:
<125 ppm
■
Resolution: <1 µV
ni.com