Figure 5. Verifying Zero Offset Test Connections
PXIe-4082
6½-Digit DMM
4
5
1
2
3
1. SA1000X
2. Probe inputs shorting adapter
3. BNC feed-through terminator
4. Double banana (m)-to-BNC (f) adapter
5. DMM (HI and LO)
2. Connect power to the SA1000X and warm up the probe for at least 30 minutes.
3. Configure the DMM with the following settings:
■
Function: DC voltage
■
Resolution: 6.5 digits
■
Number of Averages: 4
■
Auto Zero: Enabled
■
Range: 100 mV
4. Measure the signal at the DMM.
5. Verify that the DMM reading is ±100 µV.
If the reading is outside this verification limit, return the probe to NI for repair. Refer
to
for more information on contacting NI.