2. Connect power to the SA1000X and warm up the probe for at least 30 minutes.
3. Configure the DMMs with the following settings:
Setting
DMM 1
DMM 2
Function
DC Voltage
Resolution
6.5 digits
Number of Averages
4
Auto Zero
Enabled
Range
10 V
1 V
4. Output a 5 V DC signal from the DC source.
5. Measure the 5 V signal with DMM 1; measure the output of the probe with
DMM 2.
6. Calculate the following:
Reading
DMM 1
10
− Reading
DMM 2
7. Verify that the result is <500 µV.
If the reading is outside this verification limit, return the probe to NI for repair. Refer
to
for more information on contacting NI.
Verifying Zero Offset
1. Connect the test system as follows:
ni.com
16
SA1000X Safety, Environmental, and Regulatory Information