2.
Zero the power sensor as described in the
section.
3.
Connect the RF OUT connector of the signal generator to the input port of the power
splitter using an SMA (f)-to-N (m) adapter and an SMA (m)-to-SMA (m) cable.
4.
Select one of the power splitter output ports. Refer to this power splitter output port as
splitter output 1
.
5.
Connect the 50 Ω SMA terminator (m) to
splitter output 1
.
6.
Refer to the other power splitter output as
splitter output 2
.
7.
Connect the power sensor to
splitter output 2
.
The following figure illustrates the hardware setup.
Figure 1. Connection Diagram for Measuring at Splitter 2
1
5
4
3
6
7
2
1. Signal generator
2. SMA (f)-to-N (m) adapter
3. SMA (m)-to-SMA (m) cable
4.
50 Ω
SMA terminator (m)
5. Power splitter
6. SMA (m)-to-N (f) adapter
7. Power sensor
8.
Configure the power sensor with the following settings:
•
Power measurement: continuous average
•
Path selection: automatic
•
Averaging: automatic
•
Averaging resolution: 4 (0.001 dB)
•
Aperture: 20 ms
PXIe-5172 Calibration Procedure
|
© National Instruments
|
9