13. Connect the calibrator test head to channel 2 of the PXIe-5172 using an SMB (plug) to
BNC (f) and repeat steps 2 through 8 for each configuration listed in the
14. Disable the calibrator output.
15. Connect the calibrator test head to channel 3 of the PXIe-5172 using an SMB (plug) to
BNC (f) and repeat steps 2 through 8 for each configuration listed in the
16. Disable the calibrator output.
Note
If you are verifying the PXIe-5172 (8CH), proceed to the next step. If
you are verifying the PXIe-5172 (4CH), DC Accuracy Verification is complete.
17. Connect the calibrator test head to channel 4 of the PXIe-5172 using an SMB (plug) to
BNC (f) and repeat steps 2 through 8 for each configuration listed in the
18. Disable the calibrator output.
19. Connect the calibrator test head to channel 5 of the PXIe-5172 using an SMB (plug) to
BNC (f) and repeat steps 2 through 8 for each configuration listed in the
20. Disable the calibrator output.
21. Connect the calibrator test head to channel 6 of the PXIe-5172 using an SMB (plug) to
BNC (f) and repeat steps 2 through 8 for each configuration listed in the
22. Disable the calibrator output.
23. Connect the calibrator test head to channel 7 of the PXIe-5172 using an SMB (plug) to
BNC (f) and repeat steps 2 through 8 for each configuration listed in the
24. Disable the calibrator output.
Verifying AC Amplitude Accuracy
Follow this procedure to verify the AC amplitude accuracy of the PXIe-5172 by comparing the
voltage measured by the PXIe-5172 to the voltage measured by the DMM.
Refer to the following table as you complete the following steps: