Figure 6. 1 M
Ω
Bandwidth Verification Cabling Diagram
P
XIe-5172
100 MHz Oscilloscope
50
Ω:
≤ 10V
peak
1M
Ω:
≤ 42V
peak
AUX 0
+5V MAX
CH 0
CH 1
CH 2
CH 3
CH 4
CH 5
CH 5
CH 7
7
6
9
8
4
5
1
2
3
1. Power sensor
2. SMA (m)-to-N (f) adapter
3. SMA (m)-to-SMB (plug) adapter
4. Power splitter
5.
50 Ω
SMA feed-thru terminator
6. PXIe-5172
7. SMA (m)-to-SMA (m) cable
8. SMA (f)-to-N (m) adapter
9. Signal generator
2.
Configure the power sensor with the following settings:
•
Power measurement: Continuous average
•
Path selection: automatic
•
Averaging: automatic
28
|
ni.com
|
PXIe-5172 Calibration Procedure