Table 1. PXIe-5172 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
Power
sensor
Rohde & Schwarz
NRP-Z91
Test system
characterization
Verifications:
•
Bandwidth
Power measurement:
•
Frequency range: 50 kHz to 101 MHz
•
Power range: -16 dBm to 10 dBm
•
VSWR: <1.11
•
Absolute accuracy
–
<0.048 dB for 50 kHz to <100 MHz
–
<0.063 dB for 100.1 MHz
•
Relative accuracy at
-4 dBm: <0.022 dB for 50 kHz to <101 MHz
Signal
generator
Rohde & Schwarz
SMA100A
Test system
characterization
Verifications:
•
Bandwidth
Sine wave generation:
•
Amplitude: -10 dBm to 16 dBm
•
Frequency: 50 kHz to 101 MHz
•
Harmonics: <-30 dBc
Power
splitter
Aeroflex/
Weinschel 1593
Test system
characterization
Verifications:
•
Bandwidth
•
Amplitude: -16 dBm to 16 dBm
•
Frequency: 50 kHz to 101 MHz
•
VSWR: <1.09
50 Ω SMA
terminator
(m)
Fairview
Microwave
ST1819
Test system
characterization
•
Amplitude: 10 dBm
•
Frequency: DC to 101 MHz
•
VSWR: <1.1
50 Ω SMB
terminator
(plug)
Fairview
Microwave
ST04B-P
Verifications:
•
RMS
noise
•
Frequency: DC to 200 MHz
•
VSWR: <1.2
SMA (m)-
to-SMA (m)
cable
Fairview
Microwave
FMC0202317-36
Test system
characterization
Verifications:
•
Bandwidth
•
Frequency: DC to 101 MHz
•
VSWR: <1.1
•
Length: ≤1 meter
•
Impedance: 50 Ω
PXIe-5172 Calibration Procedure
|
© National Instruments
|
5