1.
Place the 50 Ω SMA feed-thru terminator between the splitter output and the SMA (m)-
to-SMB (plug) adapter of
splitter output 2
. Connect
splitter output 2
of the power sensor
section to channel 0 of the PXIe-5172.
Note
The
power sensor assembly
must match the configuration used in the
section, in which the power sensor is connected to
splitter output 1
and the signal generator is connected to the input port of the
power splitter.
The following figure illustrates the hardware setup.
PXIe-5172 Calibration Procedure
|
© National Instruments
|