Table 1. PXIe-5172 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
Function
generator
NI PXI-5402/5406,
Keysight 33220A,
or Keysight
33500B
Verifications:
•
AC
amplitude
accuracy
Sine wave generation:
•
Amplitude:
–
0.1 V
pk-pk
to 5.0 V
pk-pk
into 50 Ω
–
0.1 V
pk-pk
to 20 V
pk-pk
into 1 MΩ
•
Frequency: 50 kHz
SMB (plug)
to BNC (f)
adapter
Fairview
Microwave
SM3633
Verifications:
•
Timebase
accuracy
•
DC
accuracy
Adjustment
•
Frequency: DC to 89 MHz
•
Impedance: 50 Ω
SMA Tee (f-
f-f)
Fairview
Microwave
SM4942
Verifications:
•
AC
amplitude
accuracy
Impedance: 50 Ω
Double
banana plug
to BNC (f)
Pasternack PE9008 Verifications:
•
AC
amplitude
accuracy
Impedance: 50 Ω
BNC (m)-
to-SMA (m)
cable (x2)
Fairview
Microwave
FMC0208315-36
Verifications:
•
AC
amplitude
accuracy
•
Length: ≤1 meter
•
Impedance: 50 Ω