Required Software
Calibrating the PXIe-5172 requires you to install the following software on the calibration
system:
•
Supported application development environment (ADE)—LabVIEW 2016 or later
•
Verification can be performed using either of the following:
–
NI-SCOPE 17.1 or later. The PXIe-5172 was first supported in NI-SCOPE 17.1.
–
LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes 16.2 or
later. The PXIe-5172 was first supported in LabVIEW Instrument Design Libraries
for Reconfigurable Oscilloscopes 16.2.
•
Adjustment can only be performed using:
–
LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes 16.2 or
later. The PXIe-5172 was first supported in LabVIEW Instrument Design Libraries
for Reconfigurable Oscilloscopes 16.2.
You can download all required software from
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
•
PXIe-5172 Getting Started Guide
•
NI High-Speed Digitizers Help
•
PXIe-5172 Specifications
•
NI Reconfigurable Oscilloscopes Help
Visit
for the latest versions of these documents.
Test Equipment
Refer to the following table for a list of necessary equipment and model recommendations for
calibration of the PXIe-5172.
If you do not have the recommended equipment, select a substitute calibration standard using
the specifications listed in the minimum requirements column of the table.
2
|
ni.com
|
PXIe-5172 Calibration Procedure