9.
Connect all mmWave radio heads to power.
Related Information
Interconnecting the Bidirectional MIMO (Baseband and IF) Modules
on page 35
Interconnecting the Bidirectional SISO (Baseband and IF) Modules
on page 29
Connecting mmRH-3642/3643/3652/3653 Radio Heads to a Bidirectional
System
Complete the following steps to connect mmRH-3642/3643/3652/3653 radio heads to a
bidirectional mmWave Transceiver System.
Note
Connect the modules of the mmWave Transceiver System for standard use
before connecting mmRH-3642/3643/3652/3653 radio heads.
1.
Connect to the DIGITAL I/O connector of the PXIe-3620 front panel using the joint end
of a dual-channel EPLSP cable.
2.
Connect the DIO adapter module to the DIO 1 connector
4
of the mmRH-3652/3653 back
panel using an HDMI(m)-to-mini-HDMI(m) cable.
3.
Connect the other DIO adapter module to the DIO 1 connector
4
of the mmRH-3642/3643
back panel using an HDMI(m)-to-mini-HDMI(m) cable.
4.
Connect the RX LO1 connector of the PXIe-3620 front panel to the LO IN RX connector
of the mmRH-3652/3653 back panel using an SMA(m)-to-SMA(m) cable.
5.
Connect the RX IF IN connector of the PXIe-3620 front panel to the IF OUT connector
of the mmRH-3652/3653 back panel using an SMA(m)-to-SMA(m) cable.
6.
Connect the TX LO1 connector of the PXIe-3620 front panel to the LO IN TX connector
of the mmRH-3642/3643 back panel using an SMA(m)-to-SMA(m) cable.
7.
Connect the TX IF OUT connector of the PXIe-3620 front panel to the IF IN connector of
the mmRH-3642/3643 back panel using an SMA(m)-to-SMA(m) cable.
4
You may connect the DIO adapter module to the DIO 0 connector instead of the DIO 1 connector.
The choice of connector does not affect results.
mmWave Transceiver System Getting Started Guide
|
© National Instruments
|
47