Figure 31. Connecting the mmRH-3657 to the Unidirectional SISO (Baseband and IF)
RX Chassis
NI PXIe-3620
IF-LO Module, 2 GHz BW
TX
LO1
IN
LO1
IF OUT
IF IN
IF SYNC IN
IF SYNC OUT
REF IN
REF OUT
LO2 OUT
ESD
SENSITIVE
LO2 IN
I+
I–
Q+
Q–
I+
I–
Q+
Q–
DIGITAL I/O
LVTTL
LO1
LO1
OUT
LO1
IN
LO1
OUT
RX
mmWave
PXIe-7902
CLK
IN
POR
T
2
POR
T
3
POR
T
4
POR
T
5
POR
T
0
POR
T
1
PXIe-3630
I/Q Digitizer
REF IN
+10 dBm
NOM
+13 dBm
MAX
1.5 Vp-p
DIFF MAX
1.5 Vp-p
DIFF MAX
0 Vcm
ALL PORTS 50
Ω
REF
OUT
0
1
2
3
Q+
Q–
I+
I–
ACCESS
ACTIVE
2
1
3
4
5
6
7
8
9
11
12
13
14
15
16
17
18
10
NI PXIe-1085
10 MHz REF OUT
IN
24 GB/s
Embedded Controller
RESET
TRIG
ACT/
LINK
ACT/
LINK
2
1
10/100
/1000
10/100
/1000
USER1
USER2
PWR OK/
FAULT
DRIVE
GPIB
PXIe-7902
CLK
IN
POR
T
2
POR
T
3
POR
T
4
POR
T
5
POR
T
0
POR
T
1
CONTROL
LO IN
ACCESS
Input Damage Levels:
LO: +12 dBm
IF:
–
20 dBm
IF OUT
ACTIVE
2A MAX
12 V
5
1
2
3
4
1. Mini-SAS HD(m)-to-Mini-SAS HD(m) (8.25 in.)
Cables
2. MMPX(m)-to-MMPX(m) Cables
3. Mini-SAS HD(m)-to-Mini-SAS HD(m) (8.25 in.)
Cable
4. SMA(m)-to-SMA(m) Cables
5. EPLSP Cable
d)
If you have a MIMO system, repeat steps a through c for the second PXIe-3620
module in the RX chassis and a second mmRH-3657.
2.
Connect the TX chassis to the mmRH-3647.
a)
Connect the DIGITAL I/O connector of the PXIe-3620 front panel in the TX chassis
to the CONTROL connector of the mmRH-3647 back panel using an EPLSP cable.
b)
Connect the TX LO1 connector of the PXIe-3620 front panel in the TX chassis to
the LO IN connector of the mmRH-3647 back panel using an SMA(m)-to-SMA(m)
cable.
c)
Connect the TX IF OUT connector of the PXIe-3620 front panel in the TX chassis to
the IF IN connector of the mmRH-3647 back panel using an SMA(m)-to-SMA(m)
cable.
mmWave Transceiver System Getting Started Guide
|
© National Instruments
|
43