National Instruments Corporation
5-1
Analog Output Series User Manual
5
Counters
Figure 5-1 shows a counter on the AO Series device.
Figure 5-1.
Counter Block Diagram
Counters 0 and 1 each have two inputs (source and gate), one output, and
two software registers, which are used to perform different operations.
Counter functionality is built into the DAQ-STC. For more information on
the DAQ-STC, refer to the
section of Chapter 1,
.
Counter Triggering
Counters support two different triggering actions: start and pause. Only
digital triggers can initiate these actions. For more information on digital
triggers, refer to the
Triggering with a Digital Source
section of
Chapter 10,
Start Trigger
A start trigger begins a finite or continuous pulse generation. When a
continuous generation is initiated, the pulses continue to generate until
you stop the operation in software. The specified number of pulses are
generated for finite generations unless the retriggerable attribute is used.
The retriggerable attribute causes the generation to restart on a subsequent
start trigger.
S
o
u
rce
O
u
t
G
a
te
S
oftw
a
re Regi
s
ter
s