User's Manual
of GLC001
Version 1, July 2023
- 4 -
1
th
Terms of Use Security
Magal low-speed BenchtopAir-refrigerate Centrifuge is based on current
technology and safety standards:
GB4793.1-2007 Measurement, control electrical equipment for laboratories safety
requirements part 1: General safety requirements
;
GB4793.7-2001 Measurement, control electrical equipment for laboratories safety
requirement special requirements for laboratory centrifuge
;
GB191-2000 Transportation packaging logo icon
;
GB6587.6-86 Transport-Test of Electronic Measurement Instruments
;
GB/T14710-1993 Medical electrical equipment environmental requirements and
test methods
;
Therefore, using must be in accordance with the requirements of the design. If
incorrect or inappropriate use happens as follows when using Magal low-speed
BenchtopAir-refrigerate Centrifuge, it will result in equipment damage or personal
injury
:
●
Use not in accordance with design requirements;
●
Operation and maintenance staff to use without training;
●
User changes content inappropriately without authorization of the design
;
●
Not pay attention to or understand the rules of safe use.
Any person who involving the using or maintenance of
Magal low-speed BenchtopAir-refrigerate Centrifuge
should read and understand the using way and rules of
safe-using of this manual.
Following rules must be totally enforced to avoid accidences
The “Manual” is one of the elements of device components of “Magal low-speed
BenchtopAir-refrigerate Centrifuge”, which should be consistent with the low-speed
refrigerated centrifuges together in order to be accessed by the operator.
The Magal low-speed BenchtopAir-refrigerate Centrifuge is designed for
clinical medicine, biology, chemistry, genetic engineering, immunology and so on.
The maximum separation speed can not exceed the density of the sample 1.2g/cm³;
when the sample is greater than the density of 1.2g/cm ³, the maximum rotor speed
must be reduced accordingly.
When the Magal low-speed BenchtopAir-refrigerate Centrifuge process is
running (in the separation of the sample or when the rotor rotation), to make sure that
the operator must not stay within 30 cm around the centrifuge, no hazardous
substances or items blocking vents within 30 cm around the centrifuge.
If in the use of Magal low-speed BenchtopAir-refrigerate Centrifuge, the