English
User's Manual
of GLC001
Version 1, July 2023
- 10 -
made of stainless steel inner liner;
The door cover adopts an explosion-proof structure, and there is a locking mechanism
at the front of the door cover. Only when the centrifuge is powered on and the rotor is
stopped, can the door cover of the centrifuge be opened by pressing the open key
on the control panel. Only when the centrifuge door cover is locked can the
centrifuge be started!
overspeed
When the operating speed of the centrifuge rotor exceeds the set speed of 400r/min,
the machine will sound an alarm. When the operating speed exceeds the maximum
rated speed of the rotor of 450r/min, the rotor will automatically stop running. The
door can only be opened after the rotor has completely stopped. After troubleshooting,
the machine will restart;
Imbalance
If the rotor rotates unevenly during operation, causing the shaft to shake beyond the
specified range, the machine will stop running in a timely manner and issue an alarm
prompt; Generally, the rotor load is unbalanced. After the operation is terminated,
open the door cover, and after troubleshooting, the operation can be restarted.
Emergency opening
During the operation of the rotor, if a sudden power outage or machine malfunction
occurs and the door cannot be opened by pressing the button, manual door opening
can be adopted (see "Fault Handling").
2.5 Machine placement requirements
2.5.1
This machine should be placed on a level table with sufficient rigidity and away
from vibration and impact equipment, avoiding direct exposure to heat sources and
sunlight.
2.5.2
There should be a space of 10cm to 15cm on all sides of this machine for
ventilation and heat dissipation.
2.5.3
After placement, the level should be adjusted and the four supporting feet at the
bottom of the equipment should be evenly supported on the table.
2.5.4
The applicable range of the instrument's working power supply is AC230V
50HZ/60HZ 10A.
This machine must be strictly and reliably grounded, and
the machine power grounding wire must be reliably
connected to the power grid grounding wire! During the operation of
the rotor, it is strictly prohibited to manually power off, otherwise it
may cause damage to the control circuit!