Nahita GLC001 Скачать руководство пользователя страница 13

English 

User's Manual

 of GLC001 

 

Version 1, July 2023 

- 12 - 

 

 

4.

 

Preparation before use 

5.

 

4.1  Transport and installation 

The  GLC001  low-speed centrifuge is transported in a packaging box with 

cushioning and protective materials inside. After opening the packaging box, remove 

the cushioning and protective materials inside. 

 

The net weight of the machine is about 30Kg. When 

handling, lift the equipment from both sides of the 

machine and balance the force! Please handle vertically and 

do not shake the machine! 

 

When transporting or transporting over long distances, please use specialized 

packaging boxes that are properly secured and maintained in a vertical position, and 

should be handled with care. 

 

4.2    Choose a reasonable placement location 

   The GLC001 low-speed centrifuge can only be used indoors, and its placement 

should meet the following requirements: 

When the centrifuge is running, a safe distance of 20cm must be kept around. No 

harmful Dangerous goods shall be placed within this safe distance, and relevant 

personnel shall not stay; 

The support or table for placing the centrifuge should be firm and not shake or vibrate; 

If a movable bracket or trolley is used, one with a locking device should be used. 

To ensure the safe operation of the centrifuge; 

If the centrifuge is placed next to or at the corner of a wall, in order to ensure smooth 

air circulation and good heat dissipation, please ensure that the distance between 

the rear and left and right sides of the centrifuge and the wall is not less than 10cm 

and 15cm respectively. 

Centrifuges should be placed away from windows to avoid direct exposure to heat and 

sunlight. 

After placing the centrifuge, the four supporting feet should be evenly supported on 

the table and adjusted horizontally. 

The room where the centrifuge is placed must be a constant temperature room with an 

Содержание GLC001

Страница 1: ...Low Speed Centrifuge Upgraded version safety rules For safety reasons please read the operating manual carefully before starting the operation Components subjected to heat humidity and biological cha...

Страница 2: ...ure the safety of using the machines the centrifuge may be damaged by following factors Chemical effects Environmental impacts including radiation of natural UV The corrosion and abrasion and wear of...

Страница 3: ...sories requirements Inspections and maintenance of the Magal low speed BenchtopAir refrigerate Centrifuge must be conducted at the specified time intervals Security Warning Signs Description Caution B...

Страница 4: ...n 10 4 3 Fixed placement machine 10 4 4 Connect the power supply correctly 10 5 Operating Instructions 11 5 1 Introduction to Control Panel and Display Interface 11 5 2 Power on 11 5 3 open 12 5 4 clo...

Страница 5: ...of the design Not pay attention to or understand the rules of safe use Any person who involving the using or maintenance of Magal low speed BenchtopAir refrigerate Centrifuge should read and understa...

Страница 6: ...fore centrifuge run sample separation it must be confirmed with appropriate rotor and ensure a solid installation When in the process of running centrifuges rotors turning or in the process to stop ce...

Страница 7: ...fety requirement of measurement and control of laboratories electrical equipment special requirements of laboratories electrical centrifuge GB191 2000 Icon of packaging Transportation GB6587 6 86 Tran...

Страница 8: ...n 1 July 2023 7 2 Introduction to GLC001 Low Speed Centrifuge 2 1 Introduction to the Appearance of GLC001 Low Speed Centrifuge Figure 1 Front Profile of GLC001 Low Speed Centrifuge Figure 2 Rear Appe...

Страница 9: ...stem cavity system drive system rotor system base system power supply system control system display system and alarm system 2 3 1 The door cover system includes door cover door hinge and damping gas s...

Страница 10: ...ponsible for the supply of mains electricity required for the normal operation of the machine 2 3 7 The control system includes settings for rotational speed and centrifugal force operating time selec...

Страница 11: ...rotor load is unbalanced After the operation is terminated open the door cover and after troubleshooting the operation can be restarted Emergency opening During the operation of the rotor if a sudden...

Страница 12: ...nt of our company regarding the technical specifications and purchase number of the optional auxiliary parts Table 1 Rotor type and technical parameters Rotor Capacity ml Number of tubes Max speed r m...

Страница 13: ...d meet the following requirements When the centrifuge is running a safe distance of 20cm must be kept around No harmful Dangerous goods shall be placed within this safe distance and relevant personnel...

Страница 14: ...or the power line of the centrifuge This power socket must be well grounded Confirm that the power line used by the centrifuge conforms to the safety specifications of the country and region where it...

Страница 15: ...isplay on the control panel lights up and after the self test of the machine is completed enter the main interface and now you can carry out the next operation 5 3 open Press the door open key on the...

Страница 16: ...stalling the rotor are as follows see Fig 5 and Fig 6 for schematic diagram Turn on the power switch and wait for the self inspection Press the open key to open the centrifuge door cover and confirm t...

Страница 17: ...English User s Manual of GLC001 Version 1 July 2023 16 Figure 5 schematic diagram of installing rotor body...

Страница 18: ...llowable load actual load 0 5 Nmax maximum rated speed Maximum allowable load maximum allowable load Actual load actual load Do not overload the rotor otherwise it will cause the rotor to explode and...

Страница 19: ...en the locking sleeve of the rotating head is not tightened on the rotating shaft 5 8 4 If the centrifuge needs to be operated repeatedly the locking sleeve must be checked for looseness after several...

Страница 20: ...cted parameter adjustment situation is as follows the parameter adjustment knob key can be pressed vertically downwards or the system will automatically flash three times as the default 5 9 5 Speed up...

Страница 21: ...value to zero When the time value shows zero the centrifuge will automatically stop and the speed will gradually decrease from the set value to zero the time required for the speed to decrease from th...

Страница 22: ...force is generally thousands of times the gravitational force g of the Earth and it is a unit used to measure the efficiency of various instruments in separating or settling objects The calculation of...

Страница 23: ...g and maintaining the centrifuge please turn off the power switch and unplug the power cord The regular or based on usage cleaning and maintenance tasks mainly involve the centrifuge casing inner cham...

Страница 24: ...6 1 3 3 Normally the rotor is cleaned once a week If separating salt solutions or other corrosive samples please clean them immediately after use If the sample is found splashing soaking or dripping o...

Страница 25: ...propriate amount of lubricating oil or apply it with lubricating paper 6 1 3 8 When disassembling the machine the power must be cut off first and the power cord connected to the back wall of the machi...

Страница 26: ...nated or specialized personnel Regular maintenance is carried out according to the methods specified in this user manual If the above requirements are not met i e damage caused by improper user use ou...

Страница 27: ...rgency door opening and turn it clockwise until the door cover is opened Then you can take out the separated samples from the centrifuge 7 2 Fault alarm information The following table lists the alarm...

Страница 28: ...ter control failure contact our company Speed sensor malfunction contact our company Erro3 Door cover protection Close the door cover again The door cover switch is damaged please contact our company...

Страница 29: ...ase check the power supply system of the machine 6 The voltage may be too low check if the power supply voltage meets the regulations The door won t open 1 The rotor has not stopped rotating and the d...

Страница 30: ...e start button 1 The electrical control circuit is broken replace the electrical control board The machine emits a burnt smoke odor 1 Cut off the power supply 2 Is the motor burnt out 3 Whether the el...

Страница 31: ...ife In order to accumulate data on machine operation and facilitate the diagnosis of faults or accidents it is necessary to carefully keep records of each operation one for each machine The recommende...

Страница 32: ...40 Applicable power supply voltage AC220V 50HZ 60HZ 10A Set time range 1 99 hours 1 59 minutes 1 59 seconds Maximum speed 6000rpm Maximum relative centrifugal force 5150 g Maximum capacity 400ml Risin...

Страница 33: ...er cord 1 national standard 3 Turning head hanging basket Supply according to the order contract 4 Special Hex key 1 5 20ml lubricating oil 1 6 Certificate of Conformity 1 7 Operation Maintenance and...

Отзывы: