Do not place the model/ appliance in a heater set at a temperature
any higher than 235°F (113°C), some dental materials (acrylic) will
begin to degrade at temperatures higher than 235°F (113°C).
9. With the Injection Sleeve removed from the device, turn on
the myerson
HotShotElite
. To set the Melt Temperature,
press the Scroll Key once until “SP1” reads on the display.
Use the up or down keys to adjust the Melt Temperature
to 392°F (200°C). Pressing both the up and down keys
simultaneously will return the display to the process
temperature. See Section 4 for more information about
setting the Melt Temperature.
10.
Allow the device reach the set temperature before
proceeding.
11. If using DurAcetal from a 225g bottle, measure out 8 grams
of
DurAcetal
(10ml mark on the measuring cup). If using
a Single Shot, pour the entire contents of the Single Shot
into the Injection Sleeve. Using the filler funnel, pour the
DurAcetal
into the Injection Sleeve.
INJECTION SLEEVE MUST BE AT ROOM TEMPERATURE
BEFORE FILLING
.
12. To load the device, slide the filled injection sleeve into the
end of the device. Then rotate the sleeve clockwise until the
sleeve locks into place. Place the loaded gun on the supplied
base as shown. Set a timer for 8 minutes and allow the injector
to heat undisturbed.
AP-0208