10) Press the
USER
key to
scroll the number and the
TARE
key to enter each
digit of the C-Cal value.
In this example, we’ll enter the C-
Cal value of 2503. However, it
should be noted that most C-CAL
values are 5 digits.
Do not push the
TARE
key two
times in a row until the entire
number is entered. C-Cal’s less
than 4 digits will not be accepted.
If the number is too small the
display will read “LittL” and go
back to the “C-CAL” message.
Error Correction during number
entry: If you input a wrong value,
press
ZERO
to step back one
digit and reenter.
For the rare C-CAL number that
includes a decimal point and the
X1000 annunciator, press the
POWER
key to enter the decimal
point followed by an additional
press of the POWER key to input
the X1000 factor.
11) Once the entire number
is entered (not blinking)
press the
TARE
key to
complete the entry of the
C-Cal value.
12) If the Calibration was
successful, the display
will read “PASS” followed
by “CAL’D”.
13) C-Cal is complete. Push
ZERO
. The next Cal
Setup menu item
appears.
14) Push
ZERO
to store the
calibration and return to
scale operation.
5 times
fixed
blinking
blinking
blinking
blinking
blinking
blinking
blinking
3$66
Store and return to
weight display
6WRUH
fixed
blinking
fixed
fixed
fixed
blinking
6HWYS
&$/·'
next Menu item
USER
F
USER
F
USER
F
USER
F
USER
F
USER
F
USER
F
USER
F
TARE
T
TARE
T
TARE
T
TARE
T
TARE
T
ZERO
O
sto
ZERO
O
sto
USER
F
TARE
T
38
MSI-4260 Port-A-Weigh Operator’s Manual
Figure 5-7. C-Cal Calibration Procedure (continued)
5.4 Calibration Setup Menu
The Calibration Setup Menu contains two items beyond Calibration: Standard and
Auto Zero Maintenance “Auto0”. In addition, more menus will appear that are
transferred from the Main Setup Menu when Legal-for-Trade settings are used.