XQ/G
&&$/
1) Remove the seal screw. With
a small screwdriver press the
Cal Switch behind the seal
screw.
The CAL Setup Menu
appears.
2) Press the
USER
key to
scroll to the C-Cal menu
selection.
3) Press
TARE
to Start the C-
Cal Procedure.
If this calibration is using a cleared
4260, the INITIAL CALIBRATION
procedure will start here. You must
set the cal unit, capacity, and count-
before proceding to step 4,
4) The display reads “UnLd”
(unload) indicating you should
remove all weight from the
hook.
5) Press
TARE
. The 4260 sets
the zero calibration point.
6) if the zero is in range, the scale
will display “PASS”.
7) Next the display reads
“CCAL?” to ask that you look
up the C-Cal value to be used.
8) Press
TARE
. The 4260 is
ready for numeric entry of the
C-CAL value.
9) To enter an C-Cal value
press
USER
. The display
far left digit will flash zero
indicating that a number
should be entered.
&$/
C-Cal Calibration
Procedure
blinking
3$66
No weight
on hook
Flashes
Zero
Input C-Cal value
blinking
&&D/
TARE
T
TARE
T
TARE
T
USER
F
USER
F
Remove
Hex Seal
Screw
Cal Switch
Access
Press Switch
behind the hole
with a small
Screwdriver
blinking
Calibration
37
Figure 5-6. C-Cal Calibration Procedure