13
Eco 40-55-75 manual revision 01/10
4.6.3 SUCTION
After positioning the exhauster and the filtering unit, connect
them to the Ø80 rigid tube (B, not supplied), if there is an
“A”-type installation.
Warning:
make sure that inside this tube there are no
flashes, shavings and/or other materials. CAV will not be
liable for any machine damage caused by foreign bodies
that could penetrate during the installation!
When finished, connect the Ø100mm fitting located on filter
holder cylinder to an antistatic flexible hose deriving from the
main suction line. “Peel” the copper wire and fold it inward (if
tube is grounded) or connect it to a ground or bonding lead.
Fasten the hose with its metal clamp.
It is absolutely
necessary to install a vacuum breaker valve on the
centralised suction line (if it is not already installed
inside the machine); it must be calibrated so that in the
worst conditions of use (only one operator), the current
absorbed by the motor does not exceed nameplate
limits.
4.6.3a OUTLET
Set outlet lines having a minimum diameter of 100mm. It is necessary to install, in-between silencer (C)
and outlet line, a rubber hose to allow for silencer removal and maintenance.
Important: The outlet lines
weight must not lie on silencer.
4.7 PRELIMINARY INSPECTION
Power supply, compressed air supply and Exhauster preparation stages for commissioning do not require
any special knowledge apart from those acquired after reading this manual.
Before starting up the Machine it is necessary to perform some inspections and checks in order to avoid
errors or accidents:
•
ensure that the Exhauster did not suffer any damage during assembly;
•
pay special attention when checking electrical parts, control panels, cables and air lines;
•
check that all external supplies connections are correct;
•
ensure that all mobile parts can move freely.
4.8 ADJUSTMENTS AND CHECKS
The Exhauster is tested at the Manufacturer's premises, before shipment. No further adjustment is
necessary.
When starting the Exhauster for the first time, proceed as follows:
•
set the Main switch QSl lever to ON (
switch is installed by the Customer
);
•
open compressed air supply cock; ensure that compressed air regulator reads 5 bars;
•
after being supplied, the Exhauster sets to automatic operation mode. Only the voluntary closure of
the remote starter contact can start the machine.
B
C
Содержание CA Eco 40
Страница 22: ...Eco 40 55 75 manual revision 01 10 22 Electric circuit diagram Eco 40 ...
Страница 23: ...23 Eco 40 55 75 manual revision 01 10 Eco 40 ...
Страница 24: ...Eco 40 55 75 manual revision 01 10 24 Eco 55 75 ...
Страница 25: ...25 Eco 40 55 75 manual revision 01 10 Eco 55 75 ...
Страница 26: ...Eco 40 55 75 manual revision 01 10 26 Eco 55 75 Eco 40 ...
Страница 27: ...27 Eco 40 55 75 manual revision 01 10 ...
Страница 33: ...33 Eco 40 55 75 manual revision 01 10 Date Operation description Signature ...