INE-

PHSJ-5 pH Meter 

14 

or to reset standard buffer solution group, or to change automatic distinguishing mode to manual 

distinguishing mode. User can select according to actual situation); 

If  user  adopt  their  own  standard  buffer  solution  (non-standard  buffer  solution)  to  calibrate  the 

electrode, manual calibration mode must be used. 
For example: user has one standard buffer solution, pH standard value under 25℃ is 2.704pH, 
pH standard value under 25.1℃ is 2.710pH, pH standard value under 25.2℃ is 2.720pH, they 
would be better to set the calibrating temperature at 25℃.   
When user start to calibrate, set the distinguishing mode to “MANUAL”, after the displaying is 
stable, press “ENTER button. Input the nominal pH value at current temperature. If the current 
temperature is 25.2℃, input 2.720pH and press “ENTER” button. The current calibrating data is 

saved. 

If  user  has  both  normal  standard  buffer  solution  and  his  own  standard  buffer  solution,  he  can 

operate under automatic & manual distinguishing mode separately. 

7. Saving function
   

After calibration, if user want to store the present measuring data, press “SAVE” at the starting 
working state. The meter springs out a window and ask “

Save current result”

.   

 

Press “ENTER” button to store;

 

Press “CANCEL” button to cancel.

The meter returns to starting working state. 

8. Print/communication function

Only  one  socket  or  one  mode  can  be  selected  when  user  is  going  to  print  or  communicate  with 

RS-232 or USB socket. User can not use two sockets together (RS-232 and USB) at the same time.     

8.1 Immediate Print 

        Print present measurement data. Connect TP-16 printer with the meter in starting working state. 

Press “PRINT” button to print. If user has stored measuring data, please select “PRINT current 
result?” to press “Enter” button. 

8.2 Immediate communication 

      If  user  want  to  communicate  with  outside  through  computer,  please  connect  RS-232  interface 

with computer  and use  << DATA connecting system >> software which is free of charge, and 

will be delivered to user together with the meter. 

8.3 Print stored data 

 

Print stored measured data. 

    Connect TP-16 printer with the meter, and set properly. At starting working state, press “PRINT” 

button  to  select  “Print  all”, then  press “ENTER”  button  to  print  out  all  information  including 
“meter name, printing data, stored number, potential value, pH value, temperature value, slope 

value, stored time. etc. 

8.4 Sending all data 

  Connect  the  meter  with  computer  properly,  operate  <<  DATA  connecting  system  >>  on  the 

computer, press “Print all” to send out. 

Содержание INE-PHSJ-5

Страница 1: ...INE PHSJ 5 pH Meter Instruction Manual 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 12 18...

Страница 2: ...eter 1 ADVICE Read this instruction in detail before operating the meter The meter should be re tested by metrological department when it is used over one year And the meter can t be used until it is...

Страница 3: ...INE PHSJ 5 pH Meter 2 CONTENT 1 GENERAL 2 SPECIFICATIONS 3 CONSTRUCTION 4 OPERATION 5 MAINTENANCE AND SERVICE 6 THE COMPLETE SETS OF THE METER 7 APPENDIX...

Страница 4: ...can be connected with TP 16 printer to print out measuring results or be connected with computer to communicate with outside Protection from electricity cutting off The stored data and parameters wil...

Страница 5: ...ng terminal 6 measuring electrode socket 7 temperature sensor socket 8 USB interface 9 RS 232 interface 10 power switch 11 power supply socket with fuse 3 Accessory of Meter 11 Q9 short circuit plug 1...

Страница 6: ...ap off the fuse socket put 0 5A 220V fuse into it and then twist the fuse cap on Connect power line with the meter Note In general the fuse has been placed in before ex factory If user matched meter w...

Страница 7: ...button SAVE button PRINT button and DEL button to enter corresponding working state directly 4 Set function User should check the system of the meter before using including system time types of temper...

Страница 8: ...er to dia4 Diagram 4 Diagram 4 Press MONTH button repeatedly to select month you need press SETUP button to enter or use entering method press button spring out a entering window enter your needed mon...

Страница 9: ...TUP button select item Default Temp press ENTER button to enter manual setting temperature Press SETUP button or direct press it once in the area of manual temperature setting spring out an input wind...

Страница 10: ...electrode at 0 is not exactly 1000 At this time in order to ensure measuring precision it is neccessary to reset R0 value according to nominal resistance value Press SETUP R0 key and input R0 value o...

Страница 11: ...standard buffer solution group 5 2 Setting standard buffer solution group The meter has automatic distinguishing function which can distinguish 10 different standard buffer solutions Multi point not...

Страница 12: ...avoid overlapping of pH value user should select corresponding standard buffer solution to calibrate User should delete all the other selection of solution which are not used When selection finished p...

Страница 13: ...e slope set as 100 potential value and temperature value In the middle of the screen it shows the current standard buffer solution group set by user At the bottom of the screen it shows the result of...

Страница 14: ...iagram 12 if user need to calibrate some other standard buffer solution please press CONTINUE button then repeat above procedure till calibration finished Diagram 12 During the calibration user can fi...

Страница 15: ...springs out a window and ask Save current result Press ENTER button to store Press CANCEL button to cancel The meter returns to starting working state 8 Print communication function Only one socket or...

Страница 16: ...delete a certain stored data At VIEW Stored Datum mode press DEL button to select Delete current one press ENTER button to delete Press CANCEL button to exit the operation Note You should be very care...

Страница 17: ...9 4 060 6 833 55 1 715 3 554 4 075 6 834 60 1 723 3 560 4 091 6 836 70 1 743 3 580 4 126 6 845 80 1 766 3 609 4 164 6 859 90 1 792 3 650 4 205 6 877 95 1 806 3 674 4 227 6 886 CLASS DENANARE 1th GRADE...

Страница 18: ...E KHC4H4O6 25 SATURATION 7 3 776 1 GRADE KH2C6H5O7 0 05 11 41 4 008 1 GRADE KHC8H4O4 0 05 10 12 6 865 1 GRADE KH2PO4 0 025 3 387 Na2HPO4 0 025 3 533 7 00 DENANARE KH2PO4 0 02 2 780 Na2HPO4 0 0275 3 88...

Страница 19: ...t 2 E 201D PH Combination Electrode 1 3 T 820D Temperature Sensor 1 4 Buffer Solution 4 7 10 5 bags for each 5 Multifunction Electrode Support Stand 1 6 Fuse 0 5A 5 7 AF30 Stylus 1 8 TP 16 Print Conne...

Отзывы: