background image

INE-

PHSJ-5 pH Meter 

If the meter is connected with Pt1000 temperature electrode, user should set temperature electrode to 

be Pt1000, otherwise, set temperature electrode to be RT2252.   

 

Direct  press  “Model  Pt1000”  or  “Model  RT2252”  on  screen,  and  press  “ENTER”  button  to

confirm.   

The  meter  sets  temperature  electrode  and  to  be  back  to  starting  working  state.  Press  “CANCEL” 
button or “X” to exit temperature electrode setting menu, the meter is back to starting working state. 

If selected temperature electrode is “Pt1000”, the meter displays “SETUP RO” button in menu, press 

this button to set resistance of Pt1000 temperature electrode at 0℃. Generally, the resistance value of 
Pt1000 temperature electrode under 0℃ is 1000Ω, considering the technics and material used, the 

resistance value of Pt1000 temperature electrode at 0℃ is not exactly 1000Ω. At this time, in order 

to  ensure  measuring  precision,  it  is  neccessary  to  reset  R0  value  according  to  nominal  resistance 
value. Press „SETUP R0‟ key and input R0 value of actual nominal resistance.   

The measuring result will be different everytime when the data is amended. If you are not able to get 

R0 resistance value in using Pt1000 electrode at 0℃, please do not amend the data at will. 

4.6 Calibrating mV zero point 

To ensure high precision, please calibrate mV zero point  only after  you switch on

 

the meter in 

half an hour

 

 

Press  “SETUP”  button,  select  “Zero(mV)”,  press  “ENTER”  button  to  enter  mV  calibration
mode.  The  meter  displays  “Adjust  zero(mV),  Enter?”.  Press  “ENTER”  button  to  confirm, 
otherwise, press “CANCEL” button to be back to starting working state. 

5. View function

Viewing previous calibration data

standard buffer solution group and saved measuring result. 

Press “VIEW” button at starting working state to enter viewing mode. Please refer to Diagram 7:

 

Diagram 7 

Содержание INE-PHSJ-5

Страница 1: ...INE PHSJ 5 pH Meter Instruction Manual 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 12 18...

Страница 2: ...eter 1 ADVICE Read this instruction in detail before operating the meter The meter should be re tested by metrological department when it is used over one year And the meter can t be used until it is...

Страница 3: ...INE PHSJ 5 pH Meter 2 CONTENT 1 GENERAL 2 SPECIFICATIONS 3 CONSTRUCTION 4 OPERATION 5 MAINTENANCE AND SERVICE 6 THE COMPLETE SETS OF THE METER 7 APPENDIX...

Страница 4: ...can be connected with TP 16 printer to print out measuring results or be connected with computer to communicate with outside Protection from electricity cutting off The stored data and parameters wil...

Страница 5: ...ng terminal 6 measuring electrode socket 7 temperature sensor socket 8 USB interface 9 RS 232 interface 10 power switch 11 power supply socket with fuse 3 Accessory of Meter 11 Q9 short circuit plug 1...

Страница 6: ...ap off the fuse socket put 0 5A 220V fuse into it and then twist the fuse cap on Connect power line with the meter Note In general the fuse has been placed in before ex factory If user matched meter w...

Страница 7: ...button SAVE button PRINT button and DEL button to enter corresponding working state directly 4 Set function User should check the system of the meter before using including system time types of temper...

Страница 8: ...er to dia4 Diagram 4 Diagram 4 Press MONTH button repeatedly to select month you need press SETUP button to enter or use entering method press button spring out a entering window enter your needed mon...

Страница 9: ...TUP button select item Default Temp press ENTER button to enter manual setting temperature Press SETUP button or direct press it once in the area of manual temperature setting spring out an input wind...

Страница 10: ...electrode at 0 is not exactly 1000 At this time in order to ensure measuring precision it is neccessary to reset R0 value according to nominal resistance value Press SETUP R0 key and input R0 value o...

Страница 11: ...standard buffer solution group 5 2 Setting standard buffer solution group The meter has automatic distinguishing function which can distinguish 10 different standard buffer solutions Multi point not...

Страница 12: ...avoid overlapping of pH value user should select corresponding standard buffer solution to calibrate User should delete all the other selection of solution which are not used When selection finished p...

Страница 13: ...e slope set as 100 potential value and temperature value In the middle of the screen it shows the current standard buffer solution group set by user At the bottom of the screen it shows the result of...

Страница 14: ...iagram 12 if user need to calibrate some other standard buffer solution please press CONTINUE button then repeat above procedure till calibration finished Diagram 12 During the calibration user can fi...

Страница 15: ...springs out a window and ask Save current result Press ENTER button to store Press CANCEL button to cancel The meter returns to starting working state 8 Print communication function Only one socket or...

Страница 16: ...delete a certain stored data At VIEW Stored Datum mode press DEL button to select Delete current one press ENTER button to delete Press CANCEL button to exit the operation Note You should be very care...

Страница 17: ...9 4 060 6 833 55 1 715 3 554 4 075 6 834 60 1 723 3 560 4 091 6 836 70 1 743 3 580 4 126 6 845 80 1 766 3 609 4 164 6 859 90 1 792 3 650 4 205 6 877 95 1 806 3 674 4 227 6 886 CLASS DENANARE 1th GRADE...

Страница 18: ...E KHC4H4O6 25 SATURATION 7 3 776 1 GRADE KH2C6H5O7 0 05 11 41 4 008 1 GRADE KHC8H4O4 0 05 10 12 6 865 1 GRADE KH2PO4 0 025 3 387 Na2HPO4 0 025 3 533 7 00 DENANARE KH2PO4 0 02 2 780 Na2HPO4 0 0275 3 88...

Страница 19: ...t 2 E 201D PH Combination Electrode 1 3 T 820D Temperature Sensor 1 4 Buffer Solution 4 7 10 5 bags for each 5 Multifunction Electrode Support Stand 1 6 Fuse 0 5A 5 7 AF30 Stylus 1 8 TP 16 Print Conne...

Отзывы: