11
Diagram9
For example the icon of 3.68 corresponds to 3.557 pH standard buffer solution; the icon with
tick means the standard buffer solution is selected, vis-à-vis, the icon with contrastive colour
means the solution is operational. At this time, press “SELECT” or “
Clear
” button to select or
cancel current standard buffer solution; the quantity of standard buffer solution selected in the
standard buffer solution group to be shown on window.
Press the icon in correspond to a certain solution to make the solution operational, in case the
solution is not selected (without tick), press “SELECT” button to select current solution.
Vis-à-vis, select “
Clear
” to cancel
For example, if user select 3.557 pH solution, then press the icon in correspond to 3.68pH to make
the icon showing contrastive colour, press “SELECT” button, to tick off the icon which means
selected. At the same time, the meter turns the “SELECT” to
“Clear”
automatically. Press
“Clear”
button to cancel the selection or press “Select” to select it.
In order to avoid overlapping of pH value, user should select corresponding standard buffer solution
to calibrate. User should delete all the other selection
of solution which are not used.
When selection finished, press “ENTER” button or “X”(exit) button to exit, and return to view
calibration data mode.
(Please find attached 10 kinds of standard buffer solution to be distinguished by the meter.)
5.3. View calibration
In this mode of view calibration data, user can calibrate current slope of the electrode. Please refer
to chapter 6 for operation procedure
5.4. View stored data
If user need to consult the stored measuring data, press “VIEW” button in start working state and
select “STORED Datum”. Please refer to diagram10.
User can select relevant saving data according to the data shown in the diagram 10. Press “PgUp”,
“PgDn” button to review the data in the other pages.
At this time, if user need to print or delete the data, just press “PRINT” or “DEL” button.
Right side of the screen is the press button, they are “PRINT”, “DEL”, “PgUp”, “PgDn” button.
The screen displays the current saving number as well as total No. of saving; max 4 saving data
can be shown on each page, including time of saving, temperature value, slope value, electrode
potential and pH value.