09/ 2007 

 

2.  Safety note

The instrument is built and tested in accordance with

DIN VDE 0411 part 1/ EN 61010-1

and has left the factory in perfectly safe technical state.

To maintain this state and ensure safe operation of the appliance tester, the user 

must observe the notes and warnings given in these instructions at all times.

The  unit  may  be  used  only  in  power  circuits  within  the 

overvoltage  category  III  with  a  conductor  for  600 V  max.  to 

earth,  or  within  overvoltage  category  II  with  a  conductor  for 

000 V against ground. 

Remember that work on electrical components of all kinds is 

dangerous. Even low voltages of 30 V AC and 60 V DC may be 

dangerous  to human life. 

Before starting the appliance tester up, always check it as well 

as all cables and wires for signs of damage.

Should it appear that safe operation of the appliance tester is no longer possible, 

it should be shut down immediately and secured to prevent it being switched 

on accidentally.

It may be assumed that safe operation is no longer possible: 

-   if the instrument or the measuring cables show visible signs of damage, or

-   if the appliance tester no longer functions, or

-   after long periods of storage under unfavourable conditions, or

-   after being subjected to rough transport.

In order to avoid danger,

-   do not touch the bare prod tips of the measuring cables

-    insert the measurement lines in the appropriately 

designated measuring sockets on the multimeter

3.  Scope of delivery

The scope of delivery for the 

01

  comprises:

 

3.1  one 

01

 

3.2  one software 

01

 PC-Win MM 11

 

3.3  one serial data cable with USB 2.0 compatible connection

 

3.4  one safety measuring cable, red (L = 1.4 m; prod tip diameter = 4 mm)

 

3.5   one safety measuring cable, black (L=1.4 m; prod tip diameter = 4 mm)

 

3.6  one wire temperature sensor type K,

 

3.7  one adapter for wire temperature sensor,

 

3.8  one rubber protection frame,

 

3.9  one magnetic holder with adapter and strap

 

3.10 one compact protective pouch,

 

3.11   a 9 V block battery and two different fuses (fitted in unit as first equipment),

 

3.12 one operating instructions manual

Note on optional accessory:

-   Temperature probe (K-type) made of V4A tube (P.no 044121)

 

 application: insertion probe for soft-plastic materials, liquids, gas and air

 

 measuring range: - 196 °C up to 800 °C

 

 dimensions:  length  =  210 mm,  tube  length  =  120 mm,  tube  diameter  = 

3 mm, V4A

Parts subject to wear:

-   The 

01

 contains fuses as protection against overload:

 

 One fuse, rated current 1 A quick-acting (600 V), 10 kA, D = 10 mm, L = 35 

mm  (P.no  749598)  and  one  fuse,  rated  current  15 A  quick-acting  (600  V), 

100 kA, D = 10 mm, L = 38 mm (P.no 749595).

-  The 

01

 is supplied by a fitted 9 V block battery (IEC 6 LR 61).

-    The above-mentioned safety measuring cables ATL-2 (tested accessories, 

P.no 044118) correspond to CAT III 1000 V and are approved for a current 

of 10 A.

4.  Description of appliance tester

See figure 1: 

 

Appliance front face

The display  and  operator  control  elements  specified  in  Fig. 1 are  designated 

as follows:

 

 Digital display

,

 

for the measurement value, bar graph display, overranging 

display

0XOWLPHWUH01PSYQRWLFHSGI

www.motor-pump-ventilation.com

Содержание MN 11

Страница 1: ...w pomiarowych Placering av s kerhetsm tsladdar Emniyet l m tesisat n n sar lmas Aufstellung des BENNING MM 11 Standing up the BENNING MM 11 Installation du BENNING MM 11 Colocaci n del BENNING MM 11 P...

Страница 2: ...ra obr 11 M en teploty 11 ill 11 Misura di temperatura Fig 11 Temperatuurmeting Rys 11 Pomiar temperatury 11 Fig 11 Temperaturm tning Resim 11 Is l m Bild 12 Batteriewechsel Fig 12 Battery replacement...

Страница 3: ...hsel Fig 13 Fuse replacement Fig 13 Remplacement des fusibles Fig 13 Cambio de fusible obr 13 V m na pojistek 13 ill 13 Sostituzione fusibile Fig 13 Vervanging van de smeltzekeringen Rys 13 Wymiana be...

Страница 4: ...l m Bild 5 Wechselstrommessung Fig 5 AC current measurement Fig 5 Mesure de courant alternatif Fig 5 Medici n de corriente alterna obr 5 M en st dav ho proudu 5 AC ill 5 Misura corrente alternata Fig...

Страница 5: ...iar napi cia przemiennego 3 Fig 3 V xelsp nningsm tning Resim 3 Alternatif Gerilim l m Bild 4 Gleichstrommessung Fig 4 DC current measurement Fig 4 Mesure de courant continu Fig 4 Medici n de corrient...

Страница 6: ...ngstest med summer Resim 8 Sesli Uyar c ile S reklilik Kontrol Bild 9 Kapazit tsmessung Fig 9 Capacity Testing Fig 9 Mesure de capacit Fig 9 Medici n de capacidad obr 9 M en kapacity 9 ill 9 Misura di...

Страница 7: ...not be used in power circuits with a nominal voltage higher than 1000 V DC and 750 V AC More details in Section 6 Environmental condi tions The following symbols are used in these operating instructi...

Страница 8: ...3 Scope of delivery The scope of delivery for the comprises 3 1 one 3 2 one software 01 PC Win MM 11 3 3 one serial data cable with USB 2 0 compatible connection 3 4 one safety measuring cable red L 1...

Страница 9: ...rms is indicated by an additional flashing symbol 5 1 5 The confirms each button press with a signal sound Invalid button presses are confirmed by a double signal sound In the event of an incorrect c...

Страница 10: ...e 5 1 10 The function measured value store stores stable display values in the subdisplay In the display the symbol AH is shown simultaneously As soon as a measured value is input the subdisplay is up...

Страница 11: ...igit per second for the digital display and 20 measurements for the bar graph display 5 1 17 The 01 is switched on and off with the rotary switch Shutdown position OFF 5 1 18 The 01 switches off autom...

Страница 12: ...The RANGE button and the automatic switch off APO Auto Power Off are not functional in this mode of operation Each operation of the rotary switch interrupts the DATA LOG mode 5 3 1 LOG RATE Set scan r...

Страница 13: ...ttery should be taken out of the instrument for storage 7 Electrical specifications Note The measuring precision is specified as the sum of a relative fraction of the measured value and a number of di...

Страница 14: ...d value 40 Digit 800 mV max 200 mA 10 A 0 2 of the measured value 40 Digit 800 mV max 2 A 100 A 0 2 of the measured value 40 Digit 1 V max 10 A 1 mA 0 2 of the measured value 40 Digit 1 V max 7 4 Alte...

Страница 15: ...600 Veff The built in buzzer sounds in the case of a resistance R less than 50 Meas range Resolution Max meas current Max idling voltage 1 mV 1 1 mA 3 3 V 7 8 Capacity ranges Switch setting Largest d...

Страница 16: ...display values of 20 100 of the meas range limit value minimum impulse width 0 5 ms 0 7 30 digits for display values of 50 100 of the meas range limit value within the 2 V meas range 8 Making measure...

Страница 17: ...ect current measurement See figure 5 Alternating current measurement 8 3 Resistance measurement Using the rotary switch select the required function on the 01 If necessary use the button blue to selec...

Страница 18: ...y for the frequency measurements on the Bring the safety measuring lines into contact with the measuring points read off measured value on the digital display unit of the See figure 10 Frequency pulse...

Страница 19: ...e measuring circuit Disconnect the safety measuring cables from the Set the rotary switch to the switch setting OFF Remove the rubber protection frame from the Place the onto its front and turn the sl...

Страница 20: ...intervals by our factory service We recom mend a recalibration interval of one year Send the appliance to the following address Benning Elektrotechnik Elektronik GmbH CO KG Service Centre Robert Bosch...

Страница 21: ...t tips of the measuring cable Only grab the area appropriate for hands Insert the angled terminals in the testing or measuring device 12 Environmental notice At the end of the product s useful life pl...