background image

09/ 2007 

 

-   Place the 

01

 onto its front and turn the slotted screws of the 

battery compartment by 90 ° in an anticlockwise direction.

-   Lift the battery compartment cover off the bottom section. 

-   Lift the battery from the battery compartment and carefully disconnect the 

battery supply lines from the battery.

Do  not  undo  any  screws  on  the  printed  circuit.  Leave  the 

printed circuit in the bottom section of the housing. Danger for 

the interface LED!

-   Remove the top screw (black) and the two screws in the battery compart-

ment from the bottom section (housing bottom).

-    Now place the unit on its back again.

-  Lift the front end at its bottom section and carefully remove the same.

-   Lift one end of the defective fuse from the fuse holder.

-  Fully push the defective fuse out of the fuse holder.

-   Fit the new fuse with the same rated current, rated voltage, same discon-

nection capacity, same trigger characteristics, and same dimensions.

-   Make sure that the new fuse is seated centred in the holder. 

-   Order the battery supply lines such that they are not crushed between the 

housing parts. 

-   Engage the front end on the housing bottom and fit the three screws.

-   The battery must be connected to the battery supply lines and placed in the 

position provided for this purpose in the battery compartment.

-   Engage  the  battery  cover  into  the  bottom  section  and  turn  the  slotted 

screws of the battery cover by 90 ° in a clockwise direction.

-   Place the 

01

 into the rubber protection frame 

 

See figure 13:   

Fuse replacement

9.5  Calibration

To maintain the specified  precision of the measurement results, the instrument 

must  be  recalibrated  at  regular  intervals  by  our  factory  service.  We  recom-

mend a recalibration interval of one year. Send the appliance to the following 

address:

Benning Elektrotechnik & Elektronik GmbH & CO. KG

Service Centre

Robert-Bosch-Str. 20

D - 46397 Bocholt

9.6  Spare parts

Fuse F 1 A, 600 V, 10 kA, D = 10 mm, L = 35 mm, 

P.no. 749598

Fuse F 15 A, 600 V, 100 kA, D = 10 mm, L = 38 mm,  

P.no. 749595

0.  Application of rubber protection frame

-   You can store the safety measuring lines by winding the safety measuring 

lines around the rubber protection frame 

 and engaging the tips of the 

safety measuring lines with protection against the rubber protection frame 

  (see Figure 14).

-   You can engage a safety measuring line to the rubber protection frame 

 

such that the measuring tip protrudes freely, in order to be able to apply the 

measuring tip - jointly with the 

01

 - to a measuring point.

-   The rear support on the rubber protection frame 

 provides for an inclined 

set-up  of  the 

01

  (facilitates  reading)  or  attachment  (see 

Figure 15).

-   The rubber protection frame 

 is fitted with a lug which can be freely used 

for an attachment option.

See figure 14:   

Winding up of the safety measurement line

See figure 15:   

Assembly of the  

01

.  Technical data of the measuring accessories

 

4 mm Safety measuring cable ATL 2

-   Standard: EN 61010-031,

-   Maximum  rated  voltage  to  earth  ( )  and  measuring  category:  1000 V 

CAT III, 600 V CAT IV,

-   Maximum rated current: 10 A,

-   Protective class II (

), continuous double or reinforced insulation,

-   Contamination class: 2,

-   Length: 1.4 m, AWG 18,

-   Environmental conditions:

 

Maximum barometric elevation for making measurements: 2000 m,

 

Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %

-   Only use the measuring cables if in perfect condition and according to this 

manual, since the protection provided could otherwise be impaired.

0XOWLPHWUH01PSYQRWLFHSGI

www.motor-pump-ventilation.com

Содержание MN 11

Страница 1: ...w pomiarowych Placering av s kerhetsm tsladdar Emniyet l m tesisat n n sar lmas Aufstellung des BENNING MM 11 Standing up the BENNING MM 11 Installation du BENNING MM 11 Colocaci n del BENNING MM 11 P...

Страница 2: ...ra obr 11 M en teploty 11 ill 11 Misura di temperatura Fig 11 Temperatuurmeting Rys 11 Pomiar temperatury 11 Fig 11 Temperaturm tning Resim 11 Is l m Bild 12 Batteriewechsel Fig 12 Battery replacement...

Страница 3: ...hsel Fig 13 Fuse replacement Fig 13 Remplacement des fusibles Fig 13 Cambio de fusible obr 13 V m na pojistek 13 ill 13 Sostituzione fusibile Fig 13 Vervanging van de smeltzekeringen Rys 13 Wymiana be...

Страница 4: ...l m Bild 5 Wechselstrommessung Fig 5 AC current measurement Fig 5 Mesure de courant alternatif Fig 5 Medici n de corriente alterna obr 5 M en st dav ho proudu 5 AC ill 5 Misura corrente alternata Fig...

Страница 5: ...iar napi cia przemiennego 3 Fig 3 V xelsp nningsm tning Resim 3 Alternatif Gerilim l m Bild 4 Gleichstrommessung Fig 4 DC current measurement Fig 4 Mesure de courant continu Fig 4 Medici n de corrient...

Страница 6: ...ngstest med summer Resim 8 Sesli Uyar c ile S reklilik Kontrol Bild 9 Kapazit tsmessung Fig 9 Capacity Testing Fig 9 Mesure de capacit Fig 9 Medici n de capacidad obr 9 M en kapacity 9 ill 9 Misura di...

Страница 7: ...not be used in power circuits with a nominal voltage higher than 1000 V DC and 750 V AC More details in Section 6 Environmental condi tions The following symbols are used in these operating instructi...

Страница 8: ...3 Scope of delivery The scope of delivery for the comprises 3 1 one 3 2 one software 01 PC Win MM 11 3 3 one serial data cable with USB 2 0 compatible connection 3 4 one safety measuring cable red L 1...

Страница 9: ...rms is indicated by an additional flashing symbol 5 1 5 The confirms each button press with a signal sound Invalid button presses are confirmed by a double signal sound In the event of an incorrect c...

Страница 10: ...e 5 1 10 The function measured value store stores stable display values in the subdisplay In the display the symbol AH is shown simultaneously As soon as a measured value is input the subdisplay is up...

Страница 11: ...igit per second for the digital display and 20 measurements for the bar graph display 5 1 17 The 01 is switched on and off with the rotary switch Shutdown position OFF 5 1 18 The 01 switches off autom...

Страница 12: ...The RANGE button and the automatic switch off APO Auto Power Off are not functional in this mode of operation Each operation of the rotary switch interrupts the DATA LOG mode 5 3 1 LOG RATE Set scan r...

Страница 13: ...ttery should be taken out of the instrument for storage 7 Electrical specifications Note The measuring precision is specified as the sum of a relative fraction of the measured value and a number of di...

Страница 14: ...d value 40 Digit 800 mV max 200 mA 10 A 0 2 of the measured value 40 Digit 800 mV max 2 A 100 A 0 2 of the measured value 40 Digit 1 V max 10 A 1 mA 0 2 of the measured value 40 Digit 1 V max 7 4 Alte...

Страница 15: ...600 Veff The built in buzzer sounds in the case of a resistance R less than 50 Meas range Resolution Max meas current Max idling voltage 1 mV 1 1 mA 3 3 V 7 8 Capacity ranges Switch setting Largest d...

Страница 16: ...display values of 20 100 of the meas range limit value minimum impulse width 0 5 ms 0 7 30 digits for display values of 50 100 of the meas range limit value within the 2 V meas range 8 Making measure...

Страница 17: ...ect current measurement See figure 5 Alternating current measurement 8 3 Resistance measurement Using the rotary switch select the required function on the 01 If necessary use the button blue to selec...

Страница 18: ...y for the frequency measurements on the Bring the safety measuring lines into contact with the measuring points read off measured value on the digital display unit of the See figure 10 Frequency pulse...

Страница 19: ...e measuring circuit Disconnect the safety measuring cables from the Set the rotary switch to the switch setting OFF Remove the rubber protection frame from the Place the onto its front and turn the sl...

Страница 20: ...intervals by our factory service We recom mend a recalibration interval of one year Send the appliance to the following address Benning Elektrotechnik Elektronik GmbH CO KG Service Centre Robert Bosch...

Страница 21: ...t tips of the measuring cable Only grab the area appropriate for hands Insert the angled terminals in the testing or measuring device 12 Environmental notice At the end of the product s useful life pl...

Отзывы: