background image

09/ 2007 

 

1 - 40000 are selected by means of the button (yellow) 

 ▲ and th­e button 

 

 

▼ with­ an extended operation (2 seconds) of th­e button causing a 

fast run-through of the memory locations  (10 memory locations/ sec.). 

 

 

-   The  maximum  and  minimum  values  of  the  measuring  series  are 

retrieved by pressing the button 

 

. An extended operation 

(2 seconds) of the button 

 

 exits this mode.   

 

 

-   The  extreme  values  (MAX/  MIN)  of  the  measuring  series  are 

retrieved by pressing the button 

 

 and subsequently oper-

ating the button (yellow) 

 ▲ or th­e button 

 

 

.  Another 

press of the button 

 

 exits this mode.

 

 

 A press of the button 

  

 (DATA LOG) 

exists the DATA-LOG-

OUT mode.

 

 

 Alternatively,  the  measured  value  store  can  be  read  out  also  via  the 

supplied PC- Win MM 11 software.

 

 

 A renewed automatic save of a measuring series deletes the existing data.

6.  Environment conditions:

-   The 

01

 is intended for making measurements in dry environ-

ment.

-  Maximum barometric elevation for making measurements: 2222 m,

-   Overvoltage category/ Siting category: 

IEC 60664-1/ IEC 61010-1 → 600 V

 

category III; 1000 V category II,

-  Contamination class: 2,

-  Protection Class: IP 30 (DIN VDE 0470-1 IEC/ EN 60529)

 

 IP 30 means: Protection against access to dangerous parts and protection 

against solid impurities of a diameter > 2.5 mm, (3 - first index). No protec-

tion against water, (0 - second index).

-   Operating temperature and relative humidity: 

 

 For operating temperature from 0 °C to 30 °C: relative humidity less than 80 %

 

 For operating temperature from 31 °C to 40 °C: relative humidity less than 75 %

 

 For operating temperature from 41 °C to 50 °C: relative humidity less than 45 %

-   Storage temperature: The 

01

 can be stored at any tempera-

ture in the range from - 20 °C to + 60 °C (relative humidity from 0 to 80 %). 

The battery should be taken out of the instrument for storage.

7.  Electrical specifications

Note: The measuring precision is specified as the sum of 

 

-  a relative fraction of the measured value and 

 

-   a number of digits (counting steps of the least significant digit).

This specified measuring precision is valid for temperatures in the range from 

18 °C to 28 °C and relative humidity less than 80 %.

For the 4,000 digit counter mode, the figures with the lowest value have to be 

divided by 10!

7.  Direct voltage ranges

 (Switch setting: V DC, mV DC)

The input resistance amounts to 10 M

 (within the 400 mV range 1 G

).

Measuring range Resolution

Meas. precision

Overload protection

20 mV

1 µV

± (0.06 % of the measured value + 60 Digit)

1000 V

DC

200 mV

10 µV

± (0.06 % of the measured value + 20 Digit)

1000 V

DC

2V

100 µV

± (0.06 % of the measured value + 10 Digit)

1000 V

DC

20 V

1 mV

± (0.06 % of the measured value + 10 Digit)

1000 V

DC

200 V

10 mV

± (0.06 % of the measured value + 10 Digit)

1000 V

DC

1000 V

100 mV ± (0.06 % of the measured value + 10 Digit)

1000 V

DC

7.2  Alternating voltage ranges 

(Switch setting: V AC, mV AC)

The input resistance amounts to 10 M

 in parallel 100 pF. The measured value 

is obtained and displayed as a real effective value (TRUE RMS).

Selectable coupling type: AC or AC+DC. For the coupling AC+DC an additional 

error of 1 % + 80 Digit must be taken into account. In the case of non-sinus-

shaped curves the display value becomes less precise. Thus an additional error 

results for the following crest factors:

Crest factor from 1.4 to 3.0 additional  1.5 %

Crest factor from 3.0 to 4.0 additional  3.0 %

Measuring range Frequency range

Meas. precision 

*

1

Overload protection

20 mV, 200 mV

40 Hz ~ 100 Hz

± (0.7 % of the measured value + 80 Digit)

750 V

eff

100 Hz ~ 1 kHz

± (1.0 % of the measured value + 80 Digit)

750 V

eff

2 V, 20 V

40 Hz ~ 100 Hz

± (0.7 % of the measured value + 50 Digit)

750 V

eff

0XOWLPHWUH01PSYQRWLFHSGI

www.motor-pump-ventilation.com

Содержание MN 11

Страница 1: ...w pomiarowych Placering av s kerhetsm tsladdar Emniyet l m tesisat n n sar lmas Aufstellung des BENNING MM 11 Standing up the BENNING MM 11 Installation du BENNING MM 11 Colocaci n del BENNING MM 11 P...

Страница 2: ...ra obr 11 M en teploty 11 ill 11 Misura di temperatura Fig 11 Temperatuurmeting Rys 11 Pomiar temperatury 11 Fig 11 Temperaturm tning Resim 11 Is l m Bild 12 Batteriewechsel Fig 12 Battery replacement...

Страница 3: ...hsel Fig 13 Fuse replacement Fig 13 Remplacement des fusibles Fig 13 Cambio de fusible obr 13 V m na pojistek 13 ill 13 Sostituzione fusibile Fig 13 Vervanging van de smeltzekeringen Rys 13 Wymiana be...

Страница 4: ...l m Bild 5 Wechselstrommessung Fig 5 AC current measurement Fig 5 Mesure de courant alternatif Fig 5 Medici n de corriente alterna obr 5 M en st dav ho proudu 5 AC ill 5 Misura corrente alternata Fig...

Страница 5: ...iar napi cia przemiennego 3 Fig 3 V xelsp nningsm tning Resim 3 Alternatif Gerilim l m Bild 4 Gleichstrommessung Fig 4 DC current measurement Fig 4 Mesure de courant continu Fig 4 Medici n de corrient...

Страница 6: ...ngstest med summer Resim 8 Sesli Uyar c ile S reklilik Kontrol Bild 9 Kapazit tsmessung Fig 9 Capacity Testing Fig 9 Mesure de capacit Fig 9 Medici n de capacidad obr 9 M en kapacity 9 ill 9 Misura di...

Страница 7: ...not be used in power circuits with a nominal voltage higher than 1000 V DC and 750 V AC More details in Section 6 Environmental condi tions The following symbols are used in these operating instructi...

Страница 8: ...3 Scope of delivery The scope of delivery for the comprises 3 1 one 3 2 one software 01 PC Win MM 11 3 3 one serial data cable with USB 2 0 compatible connection 3 4 one safety measuring cable red L 1...

Страница 9: ...rms is indicated by an additional flashing symbol 5 1 5 The confirms each button press with a signal sound Invalid button presses are confirmed by a double signal sound In the event of an incorrect c...

Страница 10: ...e 5 1 10 The function measured value store stores stable display values in the subdisplay In the display the symbol AH is shown simultaneously As soon as a measured value is input the subdisplay is up...

Страница 11: ...igit per second for the digital display and 20 measurements for the bar graph display 5 1 17 The 01 is switched on and off with the rotary switch Shutdown position OFF 5 1 18 The 01 switches off autom...

Страница 12: ...The RANGE button and the automatic switch off APO Auto Power Off are not functional in this mode of operation Each operation of the rotary switch interrupts the DATA LOG mode 5 3 1 LOG RATE Set scan r...

Страница 13: ...ttery should be taken out of the instrument for storage 7 Electrical specifications Note The measuring precision is specified as the sum of a relative fraction of the measured value and a number of di...

Страница 14: ...d value 40 Digit 800 mV max 200 mA 10 A 0 2 of the measured value 40 Digit 800 mV max 2 A 100 A 0 2 of the measured value 40 Digit 1 V max 10 A 1 mA 0 2 of the measured value 40 Digit 1 V max 7 4 Alte...

Страница 15: ...600 Veff The built in buzzer sounds in the case of a resistance R less than 50 Meas range Resolution Max meas current Max idling voltage 1 mV 1 1 mA 3 3 V 7 8 Capacity ranges Switch setting Largest d...

Страница 16: ...display values of 20 100 of the meas range limit value minimum impulse width 0 5 ms 0 7 30 digits for display values of 50 100 of the meas range limit value within the 2 V meas range 8 Making measure...

Страница 17: ...ect current measurement See figure 5 Alternating current measurement 8 3 Resistance measurement Using the rotary switch select the required function on the 01 If necessary use the button blue to selec...

Страница 18: ...y for the frequency measurements on the Bring the safety measuring lines into contact with the measuring points read off measured value on the digital display unit of the See figure 10 Frequency pulse...

Страница 19: ...e measuring circuit Disconnect the safety measuring cables from the Set the rotary switch to the switch setting OFF Remove the rubber protection frame from the Place the onto its front and turn the sl...

Страница 20: ...intervals by our factory service We recom mend a recalibration interval of one year Send the appliance to the following address Benning Elektrotechnik Elektronik GmbH CO KG Service Centre Robert Bosch...

Страница 21: ...t tips of the measuring cable Only grab the area appropriate for hands Insert the angled terminals in the testing or measuring device 12 Environmental notice At the end of the product s useful life pl...

Отзывы: