Cephalo Exposures
59
7. With the Up or Down key, raise or lower the craniostat until
the ear rods line up with the patient’s outer ear orifice and
then release the key.
8. Grip the ear rod plates with both hands and carefully close
them until the ear rods go into the patient’s ears.
Be extremely careful when inserting the ear rods and do
not move the craniostat after they have been inserted.
This could seriously injure the patient.
9. Use the key for the Frankfort Plane Beam to line it up with
the patient’s Frankfort Plane and make sure that it is
perpendicular to the patient’s mid-sagittal plane.
Adjust the length of the nasion plate and put it against the
patient’s nasion.
CLASS 2 LASER PRODUCT: A class 2 laser is used for
the positioning beams.
The laser beams could damage the eyes.
Do not stare into the positioning beams. Warn the patient
not to look at the positioning beams.
Ear Rod Plate
Frankfort Plane Beam
Nasion Plate
Ear Rod