Panorama Exposures
31
Bite Block Positioning (Option)
*
Some part of the bite block will appear in the image.
1. Replace the chin rest with the bite block.
Put a cover on the bite block.
2. Have the patient put on an X-ray protection apron.
Open the temple stabilizers and have the patient step up to
the unit.
•
A new, uncontaminated bite block cover must be used
for each patient to prevent from cross-contamination.
•
Take care that the temple stabilizers do not hit the
patient in the eye.
Usage Note
♦
Keep bite block covers in a clean, uncontaminated area.
3. Use the Up and Down keys to match the lift with the height
of the patient.
Keep fingers away from gaps for moving parts such as the
cassette and the temple stabilizers as well as the holes on
the support column.
Usage Note
♦
Always use the Up and Down Keys to move the Lift. Never
try to force it; this would damage it.
♦
The over current protection circuit could be activated after
3 seconds for reasons such as an excessive load or the
failure to insert the cassette. The lift will stop and an
audible signal will sound. The unit will stop working and
one of the following messages will appear in the display
on the frame:
LIFT OVERLOAD! PROTECTOR ACTIVE
LIFT OVERTIME! PROTECTOR ACTIVE
To restore normal working order, press any one of the
occlusal type keys underneath the display on the patient
frame.
Lift Down Key
Lift Up Key