Electromagnetic Disturbances (EMD)
The Veraviewepocs 2D (hereafter "this device") conforms to IEC 60601-1-2:2014 Ed. 4.0, the relevant
international standard for electromagnetic disturbances (EMD).
The following is the “Guidance and Manufacturer's Declaration” which is required by IEC 60601-1-2:2014 Ed.
4.0, the relevant international standard for electromagnetic disturbances.
This is a Group 1, Class B product according to EN 55011 (CISPR 11).
This means that this device does not generate and/or use internationally radio-frequency energy, in the form of
electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or
inspection/analysis purpose and that it is suitable for use in domestic establishments and in establishments directly
connected to a low voltage power supply network which supplies buildings use for domestic purposes.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
Conducted disturbance
CISPR 11
Group 1
Class B
This device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Radiated disturbance
CISPR 11
Group 1
Class B
This device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic current
IEC 61000-3-2
Class A
Voltage fluctuations and
flicker
IEC 61000-3-3
Clause 5
•
The use environment of this device is the Professional healthcare facility environment.
•
This device needs special precautions regarding EMD and needs to be installed and put into service
according to the EMD information provided in the ACCOMPANYING DOCUMENTS.
•
Use of parts other than those accompanied or specified by J. MORITA MFG. CORP. could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this device and result
in improper operation.
•
Do not use this device as adjacent or stacked as possible with other.
When adjoining or stacking is necessary, use it after observing whether this equipment and other
equipment work properly.
•
Portable and mobile RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm to any part of the X550, including cables
specified by the manufacturer.