4

Operation  2018-10-21

If an accident occurs, the Root ZX mini must not be used until repairs have been completed by a 
qualified and trained technician authorized by the manufacturer.

This equipment must only be used by dentists and other legally licensed professionals.

In Case of Accident

Intended Operator Profile

Patient Population

 

J. MORITA MFG. CORP. will not be responsible for accidents, equipment damage, or bodily 
injury resulting from:

1. Repairs made by personnel not authorized by J. MORITA MFG. CORP.
2. Any changes, modifications, or alterations of its products
3. The use of products or equipment made by other manufacturers, except for those by J. 

MORITA MFG. CORP.

4. Maintenance or repairs using parts or components other than those specified by J. MORITA 

MFG. CORP. and other than in their original condition

5. Operating the equipment in ways other than the operating procedures described in this man-

ual or resulting from the safety precautions and warnings in this manual not being observed

6. Workplace conditions and environment or installation conditions which do not conform to 

those stated in this manual such as improper electrical power supply

7. Fires, earthquakes, floods, lightning, natural disasters, or acts of God.

 

The useful life of the Root ZX mini is 6 years (based on self-certification) from the date of ship-
ment provided it is regularly and properly inspected and maintained. 

 

J. MORITA MFG. CORP. will supply replacement parts and be able to repair the product for a 
period of 10 years after the manufacture of the product has been discontinued. For the duration 
of this period, we will supply replacement parts and be able to repair the product.

Disclaimer

Age

Child to Elderly 

Weight

N/A

Nationality

N/A

Sex

N/A

Health

It is not intended for use on patients wearing pacemakers or ICDs.

Condition

Conscious and mentally alert person. (Person who can stay still 
during treatment.)

• This equipment is not recommended for use in children under 12 years of age.

Содержание Root ZX mini

Страница 1: ...Thinking ahead Focused on life Apex Locator Root ZX mini Operation Instructions...

Страница 2: ...he Unit 10 Operation Panel Display and Switches 10 Settings 11 Meter Display 13 Root Canal Not Suitable for Electronic Measurement 16 Root ZX mini Meter Reading and Radiography 18 3 After Using the Un...

Страница 3: ...ecifications 30 Specifications 30 Symbols 31 Electromagnetic Disturbances EMD 32 Thank you for purchasing the Root ZX mini For optimum safety and performance read this manual thoroughly before using t...

Страница 4: ...accidents Problems and ac cidents are best avoided by foreseeing the possibility of danger and operating the equipment in accordance with the manufacturer s recommendations First thoroughly read all p...

Страница 5: ...scribed in this man ual or resulting from the safety precautions and warnings in this manual not being observed 6 Workplace conditions and environment or installation conditions which do not conform t...

Страница 6: ...oth the Root ZX mini and the other device operate normally Do not use this equipment in conjunction with an electric scalpel or on patients who have a pacemaker Blocked canals cannot be accurately mea...

Страница 7: ...t Switch Power Switch Data Output Port to transmit display data Probe Cord Strap Holes LCD Screen Accessories Standard Accessories Parts Identification Probe Cord 1 File Holder 3 Contrary Electrode 5...

Страница 8: ...stallation give the cover a light tug to confirm it is securely attached 1 Before Using the Unit 1 Slide the cover in the direction by the arrow in the illus tration and remove it from the Root ZX min...

Страница 9: ...of the Root ZX mini 1 Press the Power switch to turn the unit on The display will appear in the LCD screen The instrument turns itself off if it is not used for 10 min utes 2 Check that the probe cord...

Страница 10: ...ZX mini s performance with the tester once a week 1 Press the Power switch to turn the unit on 2 Insert the tester into the probe cord jack Check that the meter indicates within 3 bars away from above...

Страница 11: ...ng it again Operation Panel Display and Switches Sound Volume Canal Length Indicator Bars Meter Power Switch Set Switch Battery Power Indicator Flash Bar Memory Bar Information Display Off Low High Th...

Страница 12: ...en the unit is turned off is the one that will be selected when is it turned back on again 2 Set the Flash Bar The Flash Bar can be set anywhere from 2 to Apex 0 Use it as an estimate of the canal s w...

Страница 13: ...old down the Set Switch and turn the Root ZX mini on This will change the setting of the beep from Loud to off Repeat the procedure to change it from off to Soft The setting will be memorized and stay...

Страница 14: ...tails Root Canals not suit able for Electronic Measurement Always check the measurement with an X ray In some cases an accurate measurement cannot be made because of the canal shape unusual cases or p...

Страница 15: ...ormalin cresol FC or sodium hypochlorite do not get on the contrary electrode or the file holder These could cause an adverse reaction such as inflammation Always clip the file holder to the upper par...

Страница 16: ...X ray to check the results Make sure the long file holder does not prick or pierce the patient s oral mucosa 4 Press the Set Switch to select Memory 01 02 or 03 5 Insert the file up to the Flash Bar t...

Страница 17: ...leeding to stop completely Clean the inside and opening of the canal thoroughly to get rid of all blood saliva and chemical solutions and then make a measurement Broken crown If the crown is broken an...

Страница 18: ...cutting debris on the tooth Thoroughly remove all the pulp inside the canal otherwise an accurate measurement cannot be made Caries touching the gums In this case electrical leakage through the caries...

Страница 19: ...it actually is Root ZX mini Meter Reading and Radiography Sometimes the Root ZX mini meter reading and the X ray image will not correspond This does not mean that the Root ZX mini is not working prope...

Страница 20: ...isconnect the probe cord and other cords or cables Do not pull directly on the cords when connecting or disconnecting the probe and file holder Always grip the connectors to connect and disconnect cor...

Страница 21: ...the battery power indicator is flashing on and off The unit may not function properly if the battery power is low Do not reverse the plus and minus poles Never allow the spring contact to push against...

Страница 22: ...s use LR03 alkaline Oxyride or manganese dry cells Manganese dry cells will not last as long as Oxyride or alkaline dry cells Never use rechargeable nickel hydrogen or nickel cadmium batteries All the...

Страница 23: ...er or long file holder including its inside is completely dry If any water remains inside the component expel it with an air gun or another such tool Failure to do so could result in the remaining wat...

Страница 24: ...se cracking and discoloration Never wipe components with a piece of gauze that is excessively wet with Ethanol for Disinfection Ethanol 70 vol to 80 vol Do not apply or spray with any fluid Also do no...

Страница 25: ...ving Autoclaving and drying temperatures must never exceed 135 C 275 F Excess temperature could cause the components to malfunction or could cause discoloration Take the file out of the file holder or...

Страница 26: ...or J MORITA OFFICE Temperature 10 C to 45 C 14 F to 113 F Humidity 10 to 85 without condensation Atmospheric pressure 70 kPa to 106 kPa Do not expose to X rays or direct sunlight frequently or for lon...

Страница 27: ...e indicator is within 3 lines of 1 on the meter 3 Check that the Set switch changes the memory from 01 to 02 to 03 4 Check that the probe cord can be properly plugged into its jack 5 Check that the fi...

Страница 28: ...n this period any defect that is due to faulty manufacturing or material will be remedied by repair or replacement at the manufacturer or its distributor 2 Warranty repair and service In the event of...

Страница 29: ...h the oral mucosa Clean the file holder with Ethanol for Disinfection Ethanol 70 vol to 80 vol Canal Length Indica tor overreacts or is too sensitive Measurements are too short Poor accu racy Erratic...

Страница 30: ...sorption and an accurate measurement cannot be obtained Replace or clean the file holder Canal Length Indica tor does not move at all or only when the file tip is close to the apical foramen Is the ca...

Страница 31: ...se The Root ZX mini is intended to detect the apex of the root canal Operating Principle The impedance in the root canal is measured by measur ing at two frequencies and the position of the treatment...

Страница 32: ...y from rain Manufacturer CE 0197 marking Conforms with the European Directive 93 42 EEC CE marking Conforms with the European Directive 2011 65 EU EU Authorized Representative under the European Direc...

Страница 33: ...d for electromagnetic disturbances This is a Group 1 Class B product according to EN 55011 CISPR 11 This means that this device does not generate and or use internationally radio frequency energy in t...

Страница 34: ...al input output 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply lines 1...

Страница 35: ...ency range b Interference may occur in the vicinity of equipment marked with the following sym bol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply...

Страница 36: ...uite 2 05 247 Coward Street Mascot NSW 2020 Australia T 61 0 2 9667 3555 F 61 0 2 9667 3577 J MORITA CORP MIDDLE EAST 4 Tag Al Roasaa Apartment 902 Saba Pacha 21311 Alexandria Egypt T 20 0 3 58 222 94...

Отзывы: