32
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
Conducted disturbance
CISPR 11
Group 1
Class B
This device uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Radiated disturbance
CISPR 11
Group 1
Class B
This device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies build-
ings used for domestic purposes.
Harmonic current
*1
IEC 61000-3-2
Class A
Voltage fluctuations and flicker
*1
IEC 61000-3-3
Clause 5
*1: The test is not applicable since the EUT does not have AC ports.
The
Root ZX mini (hereafter “this device”) conforms to IEC 60601-1-2:2014 Ed. 4.0, the relevant
international standard for electromagnetic disturbances (EMD).
The following is the “Guidance and Manufacturer’s Declaration” which is required by IEC 60601-1-2:2014
Ed. 4.0, the relevant international standard for electromagnetic disturbances.
This is a Group 1, Class B product according to EN 55011 (CISPR 11).
This means that this device does not generate and/or use internationally radio-frequency energy,
in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of
material or inspection/analysis purpose and that it is suitable for use in domestic establishments
and in establishments directly connected to a low voltage power supply network which supplies
buildings use for domestic purposes.
Operation 2018-10-21
Electromagnetic Disturbances (EMD)
• The use environment of this device is the Home healthcare environment.
• This device needs special precautions regarding EMD and needs to be installed and put into service ac-
cording to the EMD information provided in the ACCOMPANYING DOCUMENTS.
• Use of parts other than those accompanied or specified by J. MORITA MFG. CORP. could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this device and result in
improper operation.
• Do not use this device as adjacent or stacked as possible with other. When adjoining or stacking is neces-
sary, use it after observing whether this equipment and other equipment work properly.
• Portable and mobile RF communications equipment (including peripherals such as antenna cables and ex-
ternal antennas) should be used no closer than 30 cm to any part of the RCM-7, including cables specified
by the manufacturer.