14
The Interface Solution Experts
THZ & TDZ
ONLINE MENU
1 MODEL
2 DEVICE SETUP
3 PV
4 AO
4 LRV
6 URV
DEVICE SETUP
1 PROCESS
VARIABLE
2 SETUP
3 DIAGNOSTICS &
SERVICES
4 REVIEW
PROCESS VARIABLE
PV – EGU
2 PV – % RANGE
3 PV – AO
4 TERMINAL TEMP.
SETUP
1 CONFIGURE
SENSOR
2 CONFIGURE
OPTIONS
3 CONFIGURE
RANGE
4 CONFIGURE
OUTPUT
5 CONFIGURE
DEV INFO
6 CONFIGURE
DISPLAY
CONFIGURE SENSOR
1 SENSOR TYPE
2 OUT DAMP
3 SENSOR
INFORMATION
SENSOR TYPE
1 RTD2W
2 RTD3W
3 RTD4W
4 mV
5 OHMS2W
6 OHMS3W
7 OHMS4W
8 T/C
9 T/C w/RJC
10 POTENTIOMETER
RTD
1 PT3850 100
2 PT3850 200
3 PT3850 300...
13 NI 672 120
14 CU 427 9.035
THERMOCOUPLE
1 T/C J
2 T/C K
3 T/C E
4 T/C T
5 T/C R
6 T/C S
7 T/C B
8 T/C N
9 T/C C
SENSOR INFORMATION
1 PV – LSL
2 PV – USL
3 MIN SPAN
CONFIGURE OPTIONS
1 EGU
2 FAIL MODE
3 FILTER
4 BROKEN WIRE
CONFIGURE RANGE
1 SMART RANGING
2 INPUT CAPTURE
SMART RANGING
1 PV LRV
2 PV URV
3 PV LSL
4 PV USL
5 MIN SPAN
SET THE:
1 4mA
2 20mA
3 EXIT
CONFIGURE OUTPUT
1 ANALOG OUTPUT
2 HART OUTPUT
ANALOG OUTPUT
1 ANALOG OUT
(ZERO/FULL)
2 OUTPUT TRIM
(PROCEED/CHANGE)
HART OUTPUT
1 POLL ADDR
2 BURST MODE
3 BURST OPTION
CONFIGURE DEV INFO
1 TAG
2 DESCRIPTOR
3 MESSAGE
4 DATE
5 FINAL ASSY #
6 DEVICE ID
7 H/W REVISION
CONFIGURE DISPLAY
1 MODE
(PV, Ma, TOGGLE, %)
DIAGNOSTICS &
SERVICES
1 TEST DEVICE
(STATUS)
2 SENSOR TRIM
3 LOOP TEST
REVIEW
1 DEVICE
PARAMETERS
2 DEVICE
INFORMATION
SENSOR TRIM
1 SENSOR TRIM
2 FACTORY TRIM
LOOP TEST
1 4mA
2 20mA
3 OTHER
4 END
DEVICE PARAMETERS
MODEL
# WIRES
SENSOR TYPE
FILTER
% RANGE
LRV
URV
OUT DAMPING
MIN SPAN
EGU
SENSOR S/N
LSL
USL
FAIL STATE
LIN
BROKEN WIRE
RJC
TERM TEMP.
DEVICE INFORMATION
TAG
DESCRIPTOR
MESSAGE
DATE
UNIVERSAL REV
FIELD DEV REV
SOFTWARE REV
HARDWARE REV
PHYSICAL SIGNL CODE
POLL ADDR
BURST MODE
NUM REQ PREAMS
Key
% RANGE = Output Current in % of Span
# PREAMBLES = Number of Preamble Characters
CABLE R = Wire Resistance
DAMP = Damping Value
DEV ID = Device Identification
DIAG = Diagnostic
I OUT = Output Current
LRV = Lower Range Value
MAX VALUE = Maximum Process Value
MIN VALUE = Minimum Process Value
OFFS = Offset
POLL ADDR = Polling Address
URV = Upper Range Value
WIRE NO = Number of Input Wires
Figure 11. Summarizing the THZ and TDZ Configuration Menus in a HART Communicator
D
A
C
B
E
F
H
G
Содержание TDZ
Страница 4: ......
Страница 28: ...24 The Interface Solution Experts THZ TDZ...