SPA
6
The Interface Solution Experts
Site-Programmable RTD &
Thermocouple Limit Alarm Trips
Table 1. SPA Input Accuracy
α
3750
3850
3902
3911
3916
3923
3926
3928
672
427
Ω
1000
100, 200, 300,
400, 500, 1000
100, 200, 400,
500, 1000
100, 500
100
98.129
100, 200, 470, 500
100
120
9.035
-210°C to +770°C
(-346°F to +1418°F)
-270°C to +1390°C
(-454°F to +2534°F)
-270°C to +1013°C
(-454°F to +1855.4°F)
-270°C to +407°C
(-454°F to +764.6°F)
-50°C to +1786°C
(-58°F to +3246.8°F)
-50°C to +1786°C
(-58°F to +3246.8°F)
-270°C to +1316°C
(-454°F to +2400.8°F)
+200°C to +1836°C
(+392°F to +3336.8°F)
-10 to +120mV
Range
-185°C to +540°C
(-301°F to +1004°F)
-200°C to +850°C
(-328°F to +1742°F)
-100°C to +650°C
(-148°F to +1201°F)
-200°C to +630°C
(-328°F to +1166°F)
-200°C to +510°C
(-328°F to +950°F)
-200°C to +600°C
(-328°F to +1112°F)
-200°C to +630°C
(-328°F to +1166°F)
-200°C to +850°C
(-328°F to +1742°F)
-80°C to +320°C
(-112°F to +608°F)
-50°C to +250°C
(-58°F to +482°F)
0-4000
Ω
-180°C to +760°C
(-292°F to +1400°F)
-150°C to +1370°C
(-238°F to +2498°F)
-170°C to +1000°C
(-274°F to +1832°F)
-200°C to +400°C
(-328°F to +752°F)
0°C to +1760°C
(-32°F to +3200°F)
0°C to +1760°C
(-32°F to +3200°F)
-130°C to +1300°C
(-202°F to +2372°F)
+400°C to +1820°C
(+752°F to +3308°F)
Accuracy
±0.1°C
100
Ω
: ±0.2°C; 200, 300,
& 400
Ω
: ±0.15°C; 500, &
1000
Ω
: ±0.1°C;
100
Ω
: ±0.2°C;
200, & 400
Ω
: ±0.15°C;
500, & 1000
Ω
: ±0.1°C;
100
Ω
: ±0.2°C;
500
Ω
: ±0.1°C;
±0.2°C
±0.2°C
200 to 470
Ω
±0.15°C;
500
Ω
: ±0.1°C
±0.2°C
±0.14°C
±1.6°C
±0.4
Ω
±0.25°C
±0.3°C
±0.25°C
±0.25°C
±0.5°C
±0.5°C
±0.4°C
±0.8°C
±15
µ
V
Minimum Span
2
Single, 100
Ω
Sensor: 15°C
Averaging 100
Ω
Sensors: 15°C
Differential of 100
Ω
Sensors: 30°C
Single, 200
Ω
Sensor: 10°C
Averaging 200
Ω
Sensors: 10°C
Differential of 200
Ω
Sensors: 20°C
Single, 500 or 1000
Ω
Sensor: 7.5°C
Averaging 500 or 1000
Ω
Sensors: 7.5°C
Differential of 500 or 1000
Ω
Sensors: 15°C
Single Sensor: 10°C
Averaging Sensors: 10°C
Differential of Multiple Sensors: 20°C
Single Sensor: 100°C
Averaging Sensors: 100°C
Differential of Multiple Sensors: 200°C
30
Ω
35°C
40°C
35°C
35°C
50°C
50°C
45°C
75°C
4mV
Input Type
Selection
Pt RTD
Ni RTD
Cu RTD
Ω
T/C J
T/C K
T/C E
T/C T
T/C R
T/C S
T/C N
T/C B
Voltage - mV
1. Actual
α
is 0.003750, 0.003850, etc.
2. Recommended Minimum Span. Tighter spans may result in output inaccuracies.
Range Linear Conformance Range Accuracy Minimum Span
2