Input Zero and Full Scale —
Bench Ranging
This set of menus is available only on those units
equipped with one of the AO options. Referred to as
“Standard Ranging”, it allows the user to enter the val-
ues for the analog output 0 and 100% based on actual
input. It requires the same calibration setup shown in
Figure 10.
Figure 13 shows the menu.
1. From the Main Menu “INP Z/FS” screen,
press SELECT. This will access “SAVE
ZERO”.
2. Press SELECT. The screen will flash the
input level.
3. Adjust input to the desired 0% level. The
screen will flash the changing input level.
4. Press SELECT, storing the zero setting in
SPA memory and accessing “SAVE FULL”.
5. Press SELECT again. The screen will
flash the input level.
6. Adjust input to the desired 100% level.
The screen will flash the changing input
level.
7. Press SELECT, storing the full scale
setting in SPA memory and bringing up
“EXIT INP”.
8. Press SELECT to return to the Main Menu,
or use the arrow keys to access “SAVE
ZERO” or “SAVE FULL” again.
NOTE:
SPA zero and full scale settings are non-
interactive. It is not necessary to perform this
setting more than once.
Figure 13. Bench Ranging the AO/AOZ-equipped TPRG SPA
START
INP
Z/FS
CONF
Z/FS
TRIM
OUT
SAVE
ZERO
EXIT
INP
SAVE
FULL
####
DEGX
SAVE
FULL
SAVE
FULL
SAVE
ZERO
####
DEGX
EXIT
INP
SAVE
FULL
SAVE
ZERO
TRIM
OUT
VIEW
SELECT
VIEW
SELECT
VIEW
SELECT
VIEW
SELECT
VIEW
SELECT
VIEW
SELECT
FLASHING
FLASHING
MAIN MENU