background image

SDY

The Interface Solution Experts   5

PC-Programmable Signal 

Isolator/Converter with Display

Ordering Information

Unit

SDY 

PC-Program-

mable Isolator/

Converter with 

Display

Input

PRG

 

Programmable with 

supplied Configuration 

Software 

 

CURRENT (into 20Ω)

 

Any range from –2.5mA 

to 55mA including:

 

 

0-20mA

 

4-20mA

 

0-50mA

 

10-50mA

 

 

VOLTAGE (into 1MΩ)

 

Any range from –0.5V 

to 11V including:

 

 

0-1Vdc

 

0-5Vdc

 

1-5Vdc

 

0-10Vdc

 

(recommended mini

-

mum span, 250mV)

Output

 4-20MA

 

 User     

 scaleable  

 with  

 supplied 

 

 software

Options

-IS 

Option 

(cFMus, 

ATEX 

and 

IECEx 

Intrinsi-

cally 

Safe  

approval)

Power

10-42DC

10-30DC 

for -IS 

option

Housings

BH2NG*  2-Hub, explosion-proof enclosure with two, 

 

½-inch entry ports and a glass cover

 

BH2TG*  2-Hub, explosion-proof enclosure with two, 

 

¾-inch entry ports and a glass cover

 

BH2MG*  2-Hub, explosion-proof enclosure with two, 

 

M20 x 1.5 entry ports and a glass cover

 

D2LC  2-Hub, low base, clear cover, NEMA 4X (IP66) 

enclosure 

HP Hockey puck housing and spring clips

 

DN The DN mounting will fit the TS-32 DIN rail and the 

TS-35 DIN rail.

 

FL Mounting flanges on HP suitable for relay track or 

screw mounting 

FLD Mounting flanges on HP suitable for 3½” relay track 

mounting 

 

A

 suffix indicates ANZEx (TestSafe) Ex d approval (i.e. BH2NGA).

 

 

E

 suffix indicates ATEX (ISSeP) Ex d and Ex tD approval (i.e. BH2NGE).

 

 

2-inch pipe mount standard for A or E suffix.

 

 

P

 suffix indicates enclosure comes equipped with base and U-bolts for 

 

mounting on a 2-inch pipe (i.e. BH2NGP)

To order, specify:  Unit / Input / Output / Power [Housing]

Model Number Example:  SDY / PRG / 4-20MA / 10-42DC [BH2NG]

Complete Assemblies Available

We are your One-Stop Shop!  Free yourself from 

the hassle of looking around for pieces and parts 

by ordering a complete assembly with just one 

ordering number.   

 

We carry a complete line of housings to protect 

and complement our high-quality isolators.  Get the 

quality you need and the options you require with 

the ease of just one ordering number

.

Default Factory Configuration for 

SDY

The following are the factory default configuration 

settings for your SDY unit:

Input:

 

 

4-20mA

 

Display:

 

 

4-20mA 

 

Settings:

 

 

Input Trim OFF

 

Filter:

 

 

 60Hz

 

Output: 

 

 

4-20mA 

 

Damping:

  

 

0s

Содержание SDY

Страница 1: ...a n d M o o r e R e l i a b i l i t y All product names are registered trademarks of their respective companies User s Manual SDY PC Programmable Signal Isolator Converter with Display No 235 770 00G...

Страница 2: ...tion File from an SDY 10 Setting Output Damping 10 Trimming SDY Output 10 Scaling the SDY Output 11 Setting the Display 11 Trimming the Input 12 Creating a Custom Linearization Table 12 Installation 1...

Страница 3: ...ile 2 wire loop powered PC Programmable Signal Isolator Converter that accepts current or voltage input and provides an isolated proportional 4 20mA output See Figure 1 for a list of accommodated inpu...

Страница 4: ...asured across 250 resistor Power Supply Effect 0 002 of span per 1V change Over Voltage Performance Continued Protection 48V maximum on output 48V reverse polarity protection on output Maximum Input O...

Страница 5: ...DN mounting will fit the TS 32 DIN rail and the TS 35 DIN rail FL Mounting flanges on HP suitable for relay track or screw mounting FLD Mounting flanges on HP suitable for 3 relay track mounting A suf...

Страница 6: ...nit tagging input type and range settings and output zero span trim Prior to shipment our team of skilled technicians subjects every SDY to rigorous testing Every product Moore Industries manufactures...

Страница 7: ...26 CAUTION MOORE INDUSTRIES ISOLATING CABLE PROVIDES A GALVANIC BARRIER BETWEEN THE TRANSMITTER AND THE PC CONFIGURING IT IT IS NOT NECESSARY TO USE AN ISOLATING CABLE BUT FAILURE TO DO SO MAY RESULT...

Страница 8: ...tion of the screen will be blank and the Communication indicator will show steady ON as opposed to blinking During setup the SDY tests the availability of COM Ports 1 through 4 in your system The firs...

Страница 9: ...DY Saving the SDY Configuration File to Disk This is an optional procedure You do not have to save the file to disk in order to download the file into SDY memory Use this procedure to save information...

Страница 10: ...e box to the right of the Damping sec area and type in a number between 0 and 20 032 This number is the number of seconds that the sensor s value will be averaged over Trimming SDY Output The SDY has...

Страница 11: ...u in Figure 4 and the radio button on Figure 5 both have Input selected then Display Zero will match Input Zero and Display Full will match Input Full Figure 5 Use the Display radio button to switch f...

Страница 12: ...the table fields to set values for input and output from the transmitter s intended application Use the Up Down Left and Right arrow keys on the PC keyboard to move the cursor among the cells in the t...

Страница 13: ...points but the first and last points in the curve always default to the 0 and 100 input scale input range If you decide not to use all 85 points you must change the number of points in the Points fie...

Страница 14: ...Ground Wiring Practices Moore Industries recommends the following ground wiring practices Any Moore Industries product in a metal case or housing should be grounded The protective earth conductor must...

Страница 15: ...Display Figure 12 Connecting the HP Housed SDY PS PS 1 2 3 4 SDY 10 278 MA 10 42Vdc POWER SOURCE CURRENT METER CALIBRATED 0 50mA CURRENT SOURCE I I CALIBRATED 0 10V VOLTAGE SOURCE V V PS PS 1 2 3 4 SD...

Страница 16: ...ions of the explosion proof BH housing for the SDY unit 84mm 3 30 in 118mm 4 65 in 130 mm 5 12 in 112mm 4 40 in 82 6 mm 3 25 in 64mm 2 50 in INSTRUMENT TAG 116 mm 4 56 in 27mm 1 05 in CONDUIT FITTING...

Страница 17: ...le Signal Isolator Converter with Display FRONT VIEW SIDE VIEW 61mm 2 4 in 83mm 3 25 in 62mm 2 45 in 64mm 2 50 in 66mm 2 58 in 76mm 3 00 in 43mm 1 70 in 18mm 70 in PS PS 1 2 3 4 SDY 16 203 MA Figure 1...

Страница 18: ...Ex nA IIC Ex nA IIC When installed as Division 2 equipment the SDY shall be mounted within a tool secured enclosure which meets the requirements of ANSI ISA 61010 1or C22 2 No 1010 1 and be capable of...

Страница 19: ...ons Zone 0 Ex ia IIC The SDY shall be installed in an enclosure which maintains an ingress protection rating of at least IP20 The communications port shall be programmed through Moore Industries Cable...

Страница 20: ...led and supplied with the correct power SDY transmitters begin to operate Depending upon environmental conditions they can be expected to continue to operate unattended for extended periods of time Ma...

Страница 21: ...he installation diagram for the SDY carrying the intrinsically safe option It also includes guidelines for setting up zener barriers necessary in these types of applications These diagrams must be use...

Страница 22: ...SDY 22 The Interface Solution Experts PC Programmable Signal Isolator Converter with Display...

Страница 23: ...SDY The Interface Solution Experts 23 PC Programmable Signal Isolator Converter with Display...

Страница 24: ...SDY 24 The Interface Solution Experts PC Programmable Signal Isolator Converter with Display...

Страница 25: ...ginal Buyer only and not to Buyer s customers or the users of Buyer s products unless Buyer is an engineering contractor in which case the policy shall extend to Buyer s immediate customer only This p...

Отзывы: