Troubleshooting
SpectraMax GEMINI EM Operator’s Manual
5-3
This chapter lists error codes that may occur while using the instrument, fol-
lowed by their most likely causes and remedies. Maintenance procedures are
described in Chapter 4. For problems with the SpectraMax GEMINI EM that
are not listed here, in the U.S., contact Molecular Devices Technical Services
group at 1-800-635-5577; elsewhere, call your local representative.
It is your responsibility to decontaminate the instrument,
as well as any accessories, before requesting service by Molecular
Devices representatives and before returning the instrument or any
components to Molecular Devices Corporation.
Error Codes
and
Resolutions
If a problem occurs during operation that causes an unrecoverable (fatal)
error, the instrument will stop and an error code number will be shown in the
display on the front panel. Some errors are caused by equipment malfunction
and others result from incorrect input. Some error codes are listed in Table 5.1,
along with their probable causes and remedies. If any of the following error
codes appear on the LCD display, cycle the power to the instrument (turn
power off and then on again after a few seconds). If the error message still
appears and no speci
fi
c note appears next to that error message, call Molecu-
lar Devices Technical Support: in the United States and Canada, call (800) 635-
5577; elsewhere call your local Molecular Devices representative for assis-
tance.
BIOHAZARD:
Table 5.1: Error Codes and Resolution
Error
Code
Resolution
303
Make sure drawer is closed. Turn the instrument power
switch off and then on again. If the error message still
appears, call Technical Support.
305
Make sure nothing is blocking the drawer. Open the
drawer using the open drawer button, and make sure the
microplate is
fi
rmly seated in the carriage. Turn the
instrument power switch off and then on again. If the
error message persists, call Technical Support.
323
Warning
Warning — Low lamp (light) level. Call Molecular
Devices to arrange for lamp replacement as soon as pos-
sible.
323 Fatal Fatal error — Low lamp level. Lamp does not provide
enough light to continue operation. Contact Molecular
Devices to arrange for lamp replacement.
324
PMT calibration coef
fi
cient check failure. Turn the instru-
ment power off and then on again. If the problem per-
sists, contact Molecular Devices.
Содержание SpectraMax GEMINI EM
Страница 4: ...iv SpectraMax GEMINI EM Operator s Manual...
Страница 10: ...x SpectraMax GEMINI EM Operator s Manual...
Страница 12: ...Chapter 1 1 2 SpectraMax GEMINI EM Operator s Manual...
Страница 28: ...Chapter 1 1 18 SpectraMax GEMINI EM Operator s Manual...
Страница 30: ...Chapter 2 2 2 SpectraMax GEMINI EM Operator s Manual...
Страница 34: ...Chapter 2 2 6 SpectraMax GEMINI EM Operator s Manual...
Страница 36: ...Chapter 3 3 2 SpectraMax GEMINI EM Operator s Manual...
Страница 44: ...Chapter 3 3 10 SpectraMax GEMINI EM Operator s Manual...
Страница 46: ...Chapter 4 4 2 SpectraMax GEMINI EM Operator s Manual...
Страница 52: ...Chapter 4 4 8 SpectraMax GEMINI EM Operator s Manual...
Страница 53: ...Chapter 5 Troubleshooting Error Codes and Resolutions 5 3 Opening the Drawer Manually 5 4...
Страница 54: ...Chapter 5 5 2 SpectraMax GEMINI EM Operator s Manual...
Страница 57: ...Appendix A Cables and Accessories Cables A 3 Items Available for Use with the SpectraMax GEMINI EM A 3...
Страница 58: ...Appendix A A 2 SpectraMax GEMINI EM Operator s Manual...
Страница 60: ...Appendix A A 4 SpectraMax GEMINI EM Operator s Manual...
Страница 61: ...Appendix B Common Wavelengths for Fluorescence and Luminescence Fluorescence B 3 Luminescence B 3...
Страница 62: ...Appendix B B 2 SpectraMax GEMINI EM Operator s Manual...
Страница 64: ...Appendix B B 4 SpectraMax GEMINI EM Operator s Manual...