Maintenance
0112-0127 B
127
12.
With the copper coil spring in place around the socket, carefully
align the lamp pins in the proper orientation according to the pin
pattern. Again, do not touch the lamp window.
13.
Install the lamp with a gentle rocking motion and steady
downward pressure.
14.
Replace the diamond shaped lamp retainer over the new lamp,
and use the 3/32 inch hex key to install the two screws.
15.
Tighten the two screws securely and evenly, compressing the coil
spring until the lamp is seated.
16.
Tilt up the lamp assembly and line up the lamp with the bore of
the mounting block.
17.
Slide the lamp all the way into the block, orienting the assembly
to clear the baseplate. Slide it in until it stops.
18.
Tuck the orange power cord inside, behind the assembly.
19.
Close the instrument top cover.
20.
Use the 5/32 inch hex key to firmly tighten the two vertical
mounting screws.
21.
Tighten the clamp screw snugly but not tightly. Use caution not
to over-tighten.
22.
Replace the plastic lamp cover over the opening. Be careful not
to pinch the orange power cord if it is not tucked away.
23.
Replace and install the screw.
24.
Plug in the power cord and the computer cord.
CAUTION!
Do not over-tighten the clamp screw or the lamp
could crack inside.
FlexStation_manual.book Page 127 Tuesday, June 25, 2013 10:22 AM